FDA Adverse Event Injury Summary report: N

STYLE 15 SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 12698829 · Received October 26, 2021

Report

Report Number
9617229-2021-54086
Event Type
Injury
Date Received
October 26, 2021
Date of Event
October 1, 2021
Report Date
November 24, 2021
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED ONE EXTENDED OPENING, DARK RING AND FLAT CREASES. A WEIGHT TEST OF THE DEVICE WAS VERIFIED AND THE DEVICE UNDERWEIGHT. A MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFIED ONE SHARP EDGE OPENING IN THE SHELL WITH STRESS MARKS. A DIMENSION MEASUREMENT IN THE SHELL WAS PERFORMED WHICH IDENTIFY THE THICKNESS WITHIN SPECIFICATION. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: A SHARP EDGE OPENING IN THE SHELL WITH STRESS MARKS IN THE SIDE POSTERIOR ASSESSED AS SURGICAL IMPACT OPENING.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE RUPTURE. THE DEVICE HAS BEEN REPLACED.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE RUPTURE. THE DEVICE HAS BEEN REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1596676 STYLE 15 SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 1634464

Patients

Seq Age Sex Outcome Treatment
1 40 YR Female Required Intervention