OFFSET END CAP 11X10MM .433D OFFSET
Report
- Report Number
- 0001825034-2021-02946
- Event Type
- Malfunction
- Date Received
- October 26, 2021
- Date of Event
- September 26, 2021
- Report Date
- November 17, 2021
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- HSB
- PMA / PMN Number
- K091976
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. RADIOGRAPHS WERE PROVIDED AND REVIEWED BY A HEALTH CARE PROFESSIONAL. REVIEW OF THE AVAILABLE RECORDS IDENTIFIED THE FOLLOWING: DISPLACEMENT OF THE NAIL END CAP AT THE CALCANEUS. NO FRACTURE OR OTHER ABNORMALITY IS NOTED AND ALIGNMENT IS MAINTAINED. A SMALL AMOUNT OF RADIOLUCENCY IS NOTED ALONG THE DISTAL NAIL AND END CAP. NO PRODUCT WAS RETURNED OR PICTURES PROVIDED; VISUAL AND DIMENSIONAL EVALUATIONS COULD NOT BE PERFORMED. REVIEW OF THE DEVICE HISTORY RECORD(S) IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING RELATED TO THE REPORTED EVENT. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
(B)(4). REPORT SOURCE FOREIGN: (B)(6). CONCOMITANT MEDICAL PRODUCTS: ANKLE LOCKING NAIL 12 X 210MM CAT: 14-440321 LOT: 028130. TI-DBLE LEAD CORT 5.0X28MM SCR CAT: 14-405028 LOT: 101720. TI-DBLE LEAD CORT 5.0X28MM SCR CAT: 14-405028 LOT: 907050. TI-DBLE LEAD CORT 5.0X46MM SCR CAT: 14-405046 LOT: 916480. TI-DBLE LEAD CORT 5.0X75MM SCR CAT: 14-405075 LOT: 729800R. CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL RIGHT ANKLE PROCEDURE. SUBSEQUENTLY, THE IMPLANTS DISASSEMBLED. NO REVISION PROCEDURE HAS BEEN INDICATED AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1596272 | OFFSET END CAP 11X10MM .433D OFFSET | ROD, FIXATION | HSB | ZIMMER BIOMET, INC. | 773890R |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Male | SEE H10 |