FDA Adverse Event
Malfunction
Summary report: N
PENTAX
MDR report key: 12687244
·
Received October 25, 2021
Report
- Report Number
- 9610877-2021-50135
- Event Type
- Malfunction
- Date Received
- October 25, 2021
- Date of Event
- October 14, 2021
- Report Date
- October 25, 2021
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- FAJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EB-1575K IS AVAILABLE IN THE USA WITH A 510K NUMBER K131028. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD DRIVER PCB MALFUNCTION. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD DRIVER PCB; HOWEVER, OTHER FAILURE IS NOT RELATED TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem · 1
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1591422 | PENTAX | VIDEO CYSTOSCOPE 2.0C 5.5TP 400L | FAJ | HOYA CORPORATION PENTAX TOKYO OFFICE | ECY-1575K |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |