CARDIQUIP MODULAR HEATER COOLER
Report
- Report Number
- 3007899424-2021-00032
- Event Type
- Malfunction
- Date Received
- October 22, 2021
- Date of Event
- September 21, 2021
- Report Date
- October 12, 2022
- Manufacturer
- CARDIOQUIP, LLC
- Product Code
- DWC
- PMA / PMN Number
- K102147
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- 003
Narratives
THE MANUFACTURER IS REPORTING THE FOLLOWING COMPLAINT AFTER A VOLUNTARY REVIEW OF ALL COMPLAINTS (REPORTABLE OR NOT) SINCE 2016. THIS REPORT IS BEING FILED NOW, AFTER BEING SCRUTINIZED UNDER A NEWLY REVISED RISK MATRIX, RECENTLY ADOPTED AFTER INSPECTION. THE CUSTOMER NOTIFIED CARDIOQUIP THAT A PATIENT WAS INVOLVED DURING THE EVENT ALTHOUGH NO PATIENT INFORMATION WAS COMMUNICATED. CARDIOQUIP COULD NOT REPRODUCE THE ISSUES REPORTED BY THE CUSTOMER DURING INVESTIGATION. IT WAS DETERMINED THAT THE DEVICE WAS NOT OPERATED WITH COLD WATER IN THE WATER TANK FOR AN EXTENDED PERIOD LEADING TO AN OVERTEMP ERROR. THE DEVICE WAS INSPECTED AND IS FULLY OPERATIONAL.
THIS DEVICE HAS NOT BEEN RECEIVED BY CARDIOQUIP. ONCE THE DEVICE HAS BEEN RECEIVED AND THE INVESTIGATION HAS BEEN COMPLETED, CARDIOQUIP WILL FILE SUPPLEMENTARY REPORTS.
THE DEVICE RAN NORMALLY AT 37 C, BUT WHEN OPERATED AT 42C IT WENT OVER THE SET POINT TO 43.3 C AFTER A FEW HOURS. THE HOSES WERE CHECKED FOR KINKS OR CLOGS BUT NONE WERE FOUND. IT WAS NOTICED THE MOTOR IMPELLAR/PUMP WAS LOUDER THAN USUAL.
THE DEVICE RAN NORMALLY AT 37 C , BUT WHEN OPERATED AT 42C IT WENT OVER THE SET POINT TO 43.3 C AFTER A FEW HOURS. THE HOSES WERE CHECKED FOR KINKS OR CLOGS BUT NONE WERE FOUND. IT WAS NOTICED THE MOTOR IMPELLAR/PUMP WAS LOUDER THAN USUAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1580633 | CARDIQUIP MODULAR HEATER COOLER | CARDIOPULMONARY BYPASS COOLER HEATER | DWC | CARDIOQUIP, LLC | MCH-1000(M) |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |