FDA Adverse Event Malfunction Summary report: N

PELVIC OSTEOTOME 20MM/304MM

MDR report key: 12681989 · Received October 22, 2021

Report

Report Number
2939274-2021-06248
Event Type
Malfunction
Date Received
October 22, 2021
Report Date
September 27, 2021
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HWM
UDI-DI
10886982203120
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: H3, H4, H6: PART NUMBER: 399.86. LOT NUMBER: 4000719. MANUFACTURING SITE: UNKNOWN. RELEASE TO WAREHOUSE DATE: UNKNOWN. PART/LOT COMBINATION ARE UNKNOWN AT SITE TUTTLINGEN, NO DHR REVIEW POSSIBLE. PHOTO INVESTIGATION. THE DEVICE WAS NOT RETURNED. A PHOTO-INVESTIGATION WAS PERFORMED ON THE IMAGES. UPON INSPECTING ALL THE IMAGES PROVIDED, IT IS OBSERVED THERE IS A BROWN RESIDUE ON THE PACKAGE OF THE PELVIC OSTEOTOME 20MM/304MM, HENCE CONFIRMING THE ALLEGATION. AS THE DEVICE WAS NOT RETURNED, AN AS-RECEIVED CONDITION COULD NOT BE ASSESSED, AND A DIMENSIONAL INSPECTION AND DOCUMENT/SPECIFICATION REVIEW WERE NOT COMPLETED. CONCLUSION: THE COMPLAINT CONDITION CAN BE CONFIRMED DURING PHOTO/VIDEO INVESTIGATION. DURING THE INVESTIGATION, NO PRODUCT DESIGN ISSUES OR DISCREPANCIES WERE OBSERVED. NO MANUFACTURING ISSUES WERE NOTED DURING INVESTIGATION. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. THEREFORE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTIVE ACTION IS PROPOSED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE HANDLE IS LEAVING A BROWN RESIDUE ON THE PEEL PACK AFTER STERILIZATION. THERE WAS NO PATIENT INVOLVEMENT. THE ISSUE WAS AN INTERNAL PROCESS FOR THEIR PEEL PACKING PROCEDURES. THIS REPORT IS FOR ONE (1) PELVIC OSTEOTOME 20MM/304MM. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1577060 PELVIC OSTEOTOME 20MM/304MM OSTEOTOME HWM WRIGHTS LANE SYNTHES USA PRODUCTS LLC 399.86 4000719 10886982203120

Patients

Seq Age Sex Outcome Treatment
1 Unknown