FDA Adverse Event Malfunction Summary report: N

ON-Q* PUMP WITH SELECT-A-FLOW*

MDR report key: 12678201 · Received October 21, 2021

Report

Report Number
2026095-2021-00099
Event Type
Malfunction
Date Received
October 21, 2021
Date of Event
September 25, 2021
Report Date
December 13, 2021
Manufacturer
AVANOS MEDICAL INC.
Product Code
MEB
UDI-DI
00193494134723
PMA / PMN Number
K063530
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HISTORY RECORD FOR THE REPORTED LOT NUMBER, 80401040, IN THIS COMPLAINT WAS REVIEWED AND THE MATERIAL WAS PRODUCED ACCORDING TO THE MANUFACTURING PROCEDURES AND MET THE QUALITY REQUIREMENTS. ALL INFORMATION REASONABLY KNOWN AS OF 10-DEC-2021 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL INC.. AVANOS MEDICAL INC. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL INC. COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AVANOS MEDICAL INC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.

Additional Manufacturer Narrative · 0

THE ACTUAL COMPLAINT PRODUCT WAS NOT RETURNED FOR EVALUATION. ROOT CAUSE COULD NOT BE DETERMINED. A REVIEW OF THE DEVICE HISTORY RECORD IS IN-PROGRESS. ALL INFORMATION REASONABLY KNOWN AS OF 18-OCT-2021 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. AVANOS MEDICAL INC. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL INC. COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4).

Description of Event or Problem · 0

FILL VOLUME: 550 ML. FLOW RATE: 10 ML/HR. PROCEDURE: LEFT KNEE TOTAL REPLACEMENT. CATHPLACE: LEFT KNEE REGIONAL BLOCK. DATE OF PROCEDURE: (B)(6) 2021. INFUSION START TIME: UNKNOWN. INFUSION STOP TIME: UNKNOWN. IT WAS REPORTED THAT THE PUMP WAS EXPERIENCING A FAST FLOW ISSUE AFTER A TOTAL KNEE PROCEDURE. IT WAS ALSO REPORTED THAT THE PATIENT EXPERIENCED WEAKNESS, NUMBNESS, AND TINGLING AS A RESULT OF THE PUMP INFUSING TOO QUICKLY. ADDITIONAL INFORMATION RECEIVED 29-SEP-2021 STATED DOING WELL AND SIDE EFFECTS SUBSIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1567930 ON-Q* PUMP WITH SELECT-A-FLOW* ELASTOMERIC HFR MEB AVANOS MEDICAL INC. CB004 80401040 00193494134723

Patients

Seq Age Sex Outcome Treatment
1 76 YR Female