INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2021-54342
- Event Type
- Injury
- Date Received
- October 21, 2021
- Date of Event
- February 11, 2020
- Report Date
- December 6, 2022
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- UDI-DI
- 05060191607674
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HU
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
DEVICE PHOTO EVALUATION: VISUAL ANALYSIS OF THE PHOTOGRAPHS IDENTIFIED. LOT NUMBER IN THE PATCH: 2951806 AND CATALOG: TSM310G. OBSERVED OPENING ON POSTERIOR SIDE. DEVICE ANALYSIS PERFORMED THROUGH PHOTOGRAPHS, DUE TO THE IMPOSSIBILITY TO PERFORM MICROSCOPIC ANALYSIS IT IS NOT POSSIBLE TO DETERMINE THE MOST LIKELY FAILURE MODE.
HEALTHCARE PROFESSIONAL REPORTED FOR LEFT SIDE "AN INTRA AND EXTRA-CAPSULAR RUPTURE ON LEFT SIDE." AND "SILICON LYMPHADENOPATHY ON LEFT AXILLA AND LEFT PARASTERNAL LYMPH NODE CHAINS". DEVICE HAS BEEN EXPLANTED AND REPLACED.
HEALTH EFFECT IMPACT CODE CONT: (B)(4). A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. THE REPORTED EVENT OF "SILICON LYMPHADENOPATHY" IS PHYSIOLOGICAL COMPLICATIONS AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THESE EVENTS. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.
HEALTHCARE PROFESSIONAL REPORTED FOR LEFT SIDE "AN INTRA AND EXTRA-CAPSULAR RUPTURE ON LEFT SIDE." AND "SILICON LYMPHADENOPATHY ON LEFT AXILLA AND LEFT PARASTERNAL LYMPH NODE CHAINS". DEVICE HAS BEEN EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1570159 | INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 2951806 | 05060191607674 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR | Female | Required Intervention |