FDA Adverse Event Injury Summary report: N

INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 12676954 · Received October 21, 2021

Report

Report Number
9617229-2021-54342
Event Type
Injury
Date Received
October 21, 2021
Date of Event
February 11, 2020
Report Date
December 6, 2022
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
UDI-DI
05060191607674
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
HU
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE PHOTO EVALUATION: VISUAL ANALYSIS OF THE PHOTOGRAPHS IDENTIFIED. LOT NUMBER IN THE PATCH: 2951806 AND CATALOG: TSM310G. OBSERVED OPENING ON POSTERIOR SIDE. DEVICE ANALYSIS PERFORMED THROUGH PHOTOGRAPHS, DUE TO THE IMPOSSIBILITY TO PERFORM MICROSCOPIC ANALYSIS IT IS NOT POSSIBLE TO DETERMINE THE MOST LIKELY FAILURE MODE.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED FOR LEFT SIDE "AN INTRA AND EXTRA-CAPSULAR RUPTURE ON LEFT SIDE." AND "SILICON LYMPHADENOPATHY ON LEFT AXILLA AND LEFT PARASTERNAL LYMPH NODE CHAINS". DEVICE HAS BEEN EXPLANTED AND REPLACED.

Additional Manufacturer Narrative · 1

HEALTH EFFECT IMPACT CODE CONT: (B)(4). A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. THE REPORTED EVENT OF "SILICON LYMPHADENOPATHY" IS PHYSIOLOGICAL COMPLICATIONS AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THESE EVENTS. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED FOR LEFT SIDE "AN INTRA AND EXTRA-CAPSULAR RUPTURE ON LEFT SIDE." AND "SILICON LYMPHADENOPATHY ON LEFT AXILLA AND LEFT PARASTERNAL LYMPH NODE CHAINS". DEVICE HAS BEEN EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1570159 INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 2951806 05060191607674

Patients

Seq Age Sex Outcome Treatment
1 25 YR Female Required Intervention