PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754WWS
Report
- Report Number
- 2032227-2021-208218
- Event Type
- Malfunction
- Date Received
- October 21, 2021
- Date of Event
- October 14, 2021
- Report Date
- December 12, 2021
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- UDI-DI
- 000000763000161286
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
CUSTOMER COMPLAINT NOTES, STATED NO DELIVERY ALARMS. DEVICE PASSED THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME/A33 AND EXCESSIVE NO DELIVERY TEST. NO EXCESSIVE NO DELIVERY/OCCLUSION ALARM NOTED. DEVICE HISTORY DOWNLOAD USING THDS WAS SUCCESSFUL. HOWEVER, ON THE REPORTS DATA SHOWN: (B)(6) 2021 01:00:00 BASAL SEGMENT START EVENT, (B)(6) 2021 08:26:42 SELF TEST, (B)(6) 2021 08:28:55 REWIND, (B)(6) 2021 08:30:03 MANUAL PRIME, DELIVERED 3.8 UNITS, (B)(6) 2021 08:30:23 BASAL SEGMENT START EVENT, (B)(6) 2021 08:30:45 NORMAL BOLUS, (B)(6) 2021 00:00:00 DAILY INSULIN TOTAL 30.225 UNITS. NO MOISTURE DAMAGE FOUND ON THE ELECTRONICS, MOTOR, BATTERY TUBE AND VIBRATOR ASSEMBLY NOTED. DEVICE RECEIVED WITH MINOR SCRATCHED LCD WINDOW, BROKEN BELT CLIP SLOT, STAINED END CAP STICKER AND CRACKED RESERVOIR TUBE LIP. THE TEST P-CAP/RESERVOIR DOES LOCK INTO PLACE. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD RECURRING NO DELIVERY ALARMS. CUSTOMER WAS ASSISTED WITH TROUBLESHOOT BUT THE ISSUE WAS NOT RESOLVED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR FURTHER ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1575800 | PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754WWS | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-754WWS | B8754WWSJ | 000000763000161286 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |