BD SYRINGE 60ML ST
Report
- Report Number
- 1911916-2021-01090
- Event Type
- Malfunction
- Date Received
- October 20, 2021
- Date of Event
- September 22, 2021
- Report Date
- September 30, 2021
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: IT IS REPORTED THE PLUNGER IS DAMAGED. TO AID IN THE INVESTIGATION, ONE PHOTO WAS PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. THE PHOTO SHOWS THE BOTTOM PART OF A SYRINGE WITH A ROLLED STOPPER. NO OTHER DEFECTS OR IMPERFECTIONS WERE OBSERVED. THIS DEFECT COULD OCCUR IF THE RUBBER STOPPER WAS NOT PROPERLY ASSEMBLED TO THE PLUNGER ROD. WHEN THE PLUNGER ROD-RUBBER STOPPER WAS THEN ASSEMBLED TO THE SYRINGE BARREL, THE RUBBER STOPPER BECAME ROLLED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 302829, LOT NUMBER 9182654. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. VERIFICATION OF THE RUBBER STOPPER ASSEMBLY PROCESS WAS PERFORMED. THE SETTINGS AND ALIGNMENT WAS CORRECT AND THE FLOW OF PRODUCTS WAS GOOD. TO DATE, THERE HAVE BEEN NO OTHER SIMILAR EVENTS REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND WITH THE PHOTO SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED WHEN USING THE BD SYRINGE 60ML ST, THE DEVICE EXPERIENCED A DEFECTIVE /DEFORMED STOPPER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: THE PLUNGER IS DAMAGED. IS THE PLUNGER BROKEN OR IS THE PLUNGER RUBBER (BLACK PART) DISPLACED? PLEASE CONFIRM DEVIATIONS. IT'S THE RUBBER (STOPPER) ON THE PLUNGER, IT SEEMS TO BE "MELTED".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1561485 | BD SYRINGE 60ML ST | PISTON SYRINGE | FMF | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 9182654 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |