SPECTRA WAVEWRITER
Report
- Report Number
- 3006630150-2021-05889
- Event Type
- Injury
- Date Received
- October 20, 2021
- Date of Event
- September 28, 2021
- Report Date
- December 24, 2021
- Manufacturer
- BOSTON SCIENTIFIC NEUROMODULATION
- Product Code
- LGW
- UDI-DI
- 08714729951254
- PMA / PMN Number
- P030017
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
SC-1160 SN: (B)(4). THE RETURNED IPG WAS ANALYZED, PASSED ALL TESTS PERFORMED, AND EXHIBITED NORMAL DEVICE CHARACTERISTICS. SC-2218-50 SN: (B)(4). THE RETURNED LEAD WAS ANALYZED AND VISUAL INSPECTION REVEALED THAT THE LEAD WAS CLEANLY CUT APPROXIMATELY 23.5 CM FROM THE PROXIMAL END OF THE LEAD AND THE DISTAL PORTIONS OF THE LEAD WAS NOT RETURNED. THE CLEAN-CUT DAMAGE IS A RESULT OF A TYPICAL EXPLANT PROCEDURE AND IT IS NOT CONSIDERED A FAILURE. NO OTHER ANOMALIES WERE IDENTIFIED ON THE RETURNED PORTION OF THE LEAD. WITH ALL THE AVAILABLE INFORMATION, BOSTON SCIENTIFIC CONCLUDES: THE REPORTED EVENT WAS NOT CONFIRMED.
IT WAS REPORTED THAT THE PATIENTS IPG WAS CAUSING SHOCK. NO DEVICE MALFUNCTION WAS SUSPECTED. THE PATIENT UNDERWENT AN EXPLANT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY.
DATE OF EVENT: APPROXIMATED BASED ON THE DATE THE MANUFACTURER BECAME AWARE OF THE EVENT. ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-LINEAR LEADS, UPN: M365SC2218500, MODEL: SC-2218-50, SERIAL: (B)(4), BATCH/LOT: 7072332/ 7073714.
IT WAS REPORTED THAT THE PATIENTS IPG WAS CAUSING SHOCK. NO DEVICE MALFUNCTION WAS SUSPECTED. THE PATIENT UNDERWENT AN EXPLANT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1564823 | SPECTRA WAVEWRITER | STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF | LGW | BOSTON SCIENTIFIC NEUROMODULATION | SC-1160 | 361325 | 08714729951254 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | Female | Required Intervention |