FDA Adverse Event Malfunction Summary report: N

4.9MM HEALX ADV SP BIOC ANCHOR

MDR report key: 12666751 · Received October 20, 2021

Report

Report Number
1221934-2021-03070
Event Type
Malfunction
Date Received
October 20, 2021
Date of Event
October 8, 2021
Report Date
October 8, 2021
Manufacturer
MEDOS INTERNATIONAL SARL
Product Code
MAI
UDI-DI
10886705030576
PMA / PMN Number
K191242
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H10 ADDITIONAL NARRATIVE: INVESTIGATION SUMMARY: THE COMPLAINT DEVICE IS NOT BEING RETURNED, IT WAS DISCARDED BY THE CUSTOMER, THEREFORE UNAVAILABLE FOR A PHYSICAL EVALUATION. WITH THE INFORMATION PROVIDED, AND WITHOUT THE COMPLAINT DEVICE TO EVALUATE, WE CANNOT DETERMINE A ROOT CAUSE FOR THE REPORTED FAILURE. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER (7L47973), AND NO NON-CONFORMANCE WAS IDENTIFIED. AT THIS POINT IN TIME, NO CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. HOWEVER, DEPUY SYNTHES MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD. DEVICE HISTORY LOT: AN MRE REVIEW WAS PERFORMED AND RESULTS OBTAINED AS BELOW: PRODUCT CODE: 222422. LOT NUMBER: 7L47973. ANOMALIES OR DISCREPANCIES (NON-CONFORMANCE): NONE.

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. UDI: (B)(4). TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE 4.9MM HEALX ADV SP BIOC ANCHOR DEVICE DID NOT DEPLOY UPON INSERTION INTO THE SUBSCAPULARIS TENDON. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THE DEVICE WAS BRAND NEW AND THE FIRST USE WHEN THE ISSUE OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1561789 4.9MM HEALX ADV SP BIOC ANCHOR SOFT TISSUE ANCHOR, BIOABSORBABLE MAI MEDOS INTERNATIONAL SARL 222422 7L47973 10886705030576

Patients

Seq Age Sex Outcome Treatment
1