POLY SCW DRIVER SHFT, CANNULTD
Report
- Report Number
- 1526439-2021-02226
- Event Type
- Malfunction
- Date Received
- October 20, 2021
- Date of Event
- January 1, 2021
- Report Date
- September 23, 2021
- Manufacturer
- DEPUY SPINE INC
- Product Code
- HWR
- UDI-DI
- 10705034214177
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PHYSICIAN
Narratives
EVENT OCCURRED ON AN UNKNOWN DATE IN 2021. COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING AN ANTERIOR/POSTERIOR LUMBAR FUSION PROCEDURE ONE (1) VIPER UNIVERSAL POLY DRIVER WAS BROKEN, AND TWO (2) T20 DRIVER SHAFTS WERE STRIPPED. THE SURGEON WAS PLACING UNKNOWN SCREWS IN THE PATIENT WHEN TWO (2) POLY SCREWDRIVER SHAFTS TIPS BROKE OFF. HE THEN ATTEMPTED TO USE THE VIPER2 T20 HEXLOBE DRIVER AND T20 SCREWDRIVER SHAFT WHICH BOTH STRIPPED. HE ADVANCED USING THE VIPER UNIVERSAL POLY DRIVER AND THE TIP OF THE DRIVER WAS FOUND TO BE BROKEN ON THE BACK TABLE AFTER IT WAS TAKEN OUT. THERE WERE NO FRAGMENTS GENERATED. THERE WAS NO PATIENT CONSEQUENCE OR SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THIS REPORT IS FOR ONE (1) POLY SCW DRIVER SHFT, CANNULTD. THIS IS REPORT 5 OF 5 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1565045 | POLY SCW DRIVER SHFT, CANNULTD | DRIVER, PROSTHESIS | HWR | DEPUY SPINE INC | 286720000 | 10705034214177 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | POLY SCW DRIVER SHFT, CANNULTD| T20 SCREWDRIVER SHAFT| UNKNOWN AWLS| UNKNOWN SCREWS| VIPER UNIVERSAL POLY DRIVER| VIPER2 T20 HEXLOBE DRIVER SHAF |