FDA Adverse Event Malfunction Summary report: N

EXUFLOW FLAT FULL DUCT 10MM

MDR report key: 12663770 · Received October 20, 2021

Report

Report Number
8030107-2021-00011
Event Type
Malfunction
Date Received
October 20, 2021
Report Date
October 19, 2021
Manufacturer
DEGANIA SILICONE LTD
Product Code
OTK
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

THE BATCH NUMBER IS UNKNOWN, SO WERE UNABLE TO EVALUATE THE SPECIFIC PRODUCTION REPORTS. OUR IN-PROCESS CONTROL CONTAINS TENSILE STRENGTH TEST OF EVERY BATCH OF THE DRAINS AND 100% VISUAL INSPECTION. WE CHECKED THE TENSILE STRENGTH REPORTS OF THE RECENT BATCH OF ORHUF100 (P2139871) AND THEY SHOW IN-SPEC RESULTS. THE COMPLAINT SAMPLE WAS UNAVAILABLE FOR OUR EVALUATION AND THE BATCH NUMBER IS UNKNOWN. OUR IN-PROCESS CONTROL CONTAINS TENSILE STRENGTH TEST OF EVERY BATCH (INCLUDING CONNECTION AREA WHERE THE DRAIN TORE) AND 100% VISUAL INSPECTION. NO SHARP OBJECTS ARE PRESENT ON THE WORKSTATIONS. BASED ON THE ABOVE DATA, WE CONCLUDE THAT POSSIBILITY THAT THE DRAIN TORE FOLLOWING ANY MANUFACTURING REASON IS REMOTE; DRAIN MOST PROBABLY FAILED DUE TO SOME DAMAGE IN USE. ONE OF THE PROPERTIES OF SILICONE DRAINS IS TO SNAP EASILY IN PLACE OF SLIGHT CUT OR NICK. THE COMPLAINT CONSIDERED TO BE NOT JUSTIFIED. NO CAPA IS REQUIRED IN THIS CASE.

Description of Event or Problem · 1

BROKEN AT THE JOINT OF THE DRAIN TUBE DURING USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1565260 EXUFLOW FLAT FULL DUCT 10MM EXUFLOW FLAT FULL DUCT 10MM OTK DEGANIA SILICONE LTD 10MM UNK

Patients

Seq Age Sex Outcome Treatment
1