FDA Adverse Event Malfunction Summary report: N

GLIDESCOPE CORE QUICKCONNECT CABLE

MDR report key: 12658551 · Received October 19, 2021

Report

Report Number
9615393-2021-00304
Event Type
Malfunction
Date Received
October 19, 2021
Date of Event
September 22, 2021
Report Date
September 23, 2021
Manufacturer
VERATHON MEDICAL ULC
Product Code
CCW
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE GLIDESCOPE CORE 15-INCH MONITOR AND A GLIDESCOPE CORE QUICKCONNECT CABLE WERE RETURNED TO VERATHON FOR EVALUATION. A VERATHON TECHNICAL SERVICE REPRESENTATIVE EVALUATED THE RETURNED DEVICES BUT WAS UNABLE TO CONFIRM THE REPORTED DARK IMAGE ISSUE. VERATHON HAS COMPLETED ITS INVESTIGATION OF THE REPORTED DARK IMAGE ISSUE FOR THE COMBINATION OF THE GLIDESCOPE 15-INCH MONITOR IN CONJUNCTION WITH A GLIDESCOPE BFLEX SINGLE-USE BRONCHOSCOPE IN CAPA-2021-0003. MEASUREMENTS OF THE ANALOG VIDEO SIGNAL WERE TAKEN WHEN A DARK IMAGE WAS DISPLAYED. THE ANALOG VIDEO CLOCK SIGNAL TIMING WAS OBSERVED TO BE DELAYED, RESULTING IN THE TRANSMISSION OF THE DARK IMAGE. A DESIGN REVIEW WAS CONDUCTED, THE DELAY WAS DETERMINED TO BE RESULT OF TWO RESISTOR SETS IN THE BFLEX RECEPTACLE. IMPROVING ONE OR BOTH RESISTOR SETS WAS SHOWN TO REDUCE OR ELIMINATE THE INSTANCES OF DARK IMAGE. VERIFICATION SAMPLES WERE PRODUCED WITH THE COMPONENT IMPROVEMENTS MADE ON THE BFLEX RECEPTACLE, DESIGN VERIFICATION WAS COMPLETED, AND AN ENGINEERING CHANGE ORDER (ECO) WAS RELEASED TO IMPLEMENT THE CHANGES. AS PART OF THE INVESTIGATION A POST LAUNCH RISK ASSESSMENT (PLRA) WAS PERFORMED. THE PLRA DETERMINED THAT THE OBSERVED SEVERITY AND PROBABILITY OF OCCURRENCE FOR THIS FAILURE MODE WERE IN ALIGNMENT WITH THE PREDICTED SEVERITY AND PROBABILITY OF OCCURRENCE. DURING THE PLRA, COMPLAINT DATA WAS REVIEWED. IN ALL INSTANCES THE USER COMPLETED AN UNPLUG/PLUG SEQUENCE TO RESTORE FUNCTIONALITY OR SWITCHED TO A BACKUP DEVICE TO COMPLETE THE PROCEDURE. NO ADVERSE EVENTS WERE REPORTED. BASED ON THIS REVIEW OF THE SYSTEM RISK ASSESSMENT AND THE COMPLAINT DATA NO ADDITIONAL ACTION IS REQUIRED, VERATHON WILL CONTINUE TO MONITOR FOR TRENDS. ALL FORWARD PRODUCTION INCORPORATES THIS CHANGE.

Additional Manufacturer Narrative · 1

THE DEVICE RETURN IS ANTICIPATED, HOWEVER; AT THE TIME OF THE REPORT THE DEVICE HAS NOT BEEN RECEIVED BY VERATHON. VERATHON CONTINUES TO INVESTIGATE THE REPORTED EVENT AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT DURING A PATIENT PROCEDURE, USING A GLIDESCOPE CORE QUICKCONNECT CABLE, THE IMAGE WAS INTERMITTENT AND CONTAINED DARK SPOTS. NO DELAY IN THE PROCEDURE, USE OF A BACKUP DEVICE, OR HARM TO THE PATIENT OR USER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1560527 GLIDESCOPE CORE QUICKCONNECT CABLE LARYNGOSCOPE, RIGID CCW VERATHON MEDICAL ULC 0600-0767 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Unknown