FDA Adverse Event Malfunction Summary report: N

DIVERGENCE-L¿ ANTERIOR/OBLIQUE LUMBAR FUSION SYSTEM

MDR report key: 12653043 · Received October 19, 2021

Report

Report Number
1030489-2021-01286
Event Type
Malfunction
Date Received
October 19, 2021
Date of Event
September 20, 2021
Report Date
February 3, 2022
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC
Product Code
MAX
UDI-DI
00643169524682
PMA / PMN Number
K150135
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS #704624714:PRODUCT #2151100 LOT# CT17H012 VISUAL AND OPTICAL INSPECTION CONFIRMED THE ATTACHMENT AREA HAS BEEN DAMAGED. THERE ARE NUMEROUS WITNESS MARKS ON THE BACK SIDE OF THE HANDLE. ONE OF THE EDGES HAS BEEN BROKEN OFF AND THE OTHER SIDE HAS BEEN DEFORMED AND ROUNDED OFF CHANGING THE SHAPE OF THE AREA. THIS TYPE OF DAMAGE IS FROM REPEATED IMPACT ON THE BACK SIDE OF THE HANDLE DURING USE. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH SPINAL THERAPY. IT WAS REPORTED THAT THE INSERTER HANDLE INSTRUMENT BROKE AND THE MAIN SCREW FOR THE PITUITARY RONGEUR INSTRUMENT FELL OUT DURING USE. THERE WAS NO FRAGMENT LEFT IN THE PATIENT. THERE WERE NO SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1555934 DIVERGENCE-L¿ ANTERIOR/OBLIQUE LUMBAR FUSION SYSTEM INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, MAX MEDTRONIC SOFAMOR DANEK USA, INC 2151100 CT17H012 00643169524682

Patients

Seq Age Sex Outcome Treatment
1 41 YR Male