FDA Adverse Event Malfunction Summary report: N

CALCLOCK

MDR report key: 1264587 · Received November 12, 2008

Report

Report Number
3005039508-2008-00008
Event Type
Malfunction
Date Received
November 12, 2008
Date of Event
October 21, 2008
Report Date
October 21, 2008
Manufacturer
ORTHOHELIX SURGICAL DESIGNS, INC.
Product Code
KTT
PMA / PMN Number
K061400
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

SURGEON WAS UNABLE TO PROPERLY ENGAGE LOCKING SCREW THREADS INTO PLATE TO CREATE LOCK; THEY DIDN'T CROSS-THREAD, BUT JUST KEPT SPINNING. THIS WAS THE "WORST CALC CASE" THE SURGEON HAD EVER DONE BEFORE AND HAD A VOID LARGER THAN A GOLF BALL IN THE CALCANEO'S BONE. THE SURGEON'S FIRST SCREW WAS IN THE SLOT AND SECOND INTO A LOCKING HOLE. HE THREADED THE LOCKING DRILL GUIDE INTO THE PLATE WITH NO PROBLEM, FOLLOWED BY DRILLING WITH A .062" K-WIRE, FOLLOWED BY HAND DRILLING WITH PROVIDED BIT. ONCE THE SCREW WAS PLACED IN THE DRILLED HOLE, IT'S HEAD DROPPED ALL THE WAY TO THE PLATE INDICATING THE LACK OF BONE AROUND THE SCREW. THE SURGEON COULD NOT GET THIS SCREW TO LOCK INTO THE PLATE. THE SURGEON REMOVED THIS PLATE AND TRIED A SECOND PLATE. PRIOR TO THE TECH HANDING IT TO THE SURGEON, IT WAS VERIFIED THAT LOCKING SCREWS LOCKED INTO THE 2 LOCKING HOLES PROPERLY, AND THEY DID . THE SURGEON PROCEEDED TO PLACE 2ND PLATE ON THE BONE IN THE SAME POSITION AS THE 1ST PLATE AND HAD SAME ISSUE ACHIEVING PROPER ANGLE EVEN THOUGH VERBAL AND VISUAL AID WAS BEING GIVEN TO THE SURGEON OF THE PROPER ANGLE BETWEEN SCREW AND PLATE. THE SURGEON WAS ABLE TO LOCK THE SECOND LOCKING HOLE BUT NOT THE FIRST BECAUSE HE USED NO SCREW IN THIS HOLE. THE SURGEON THOUGHT THE PATIENT WOULD BE FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CALCLOCK FRACTURE REPAIR PLATE AND SCREW KTT ORTHOHELIX SURGICAL DESIGNS, INC. CAT-001-LR L450

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention