PENTAX
Report
- Report Number
- 9610877-2021-01123
- Event Type
- Malfunction
- Date Received
- October 13, 2021
- Date of Event
- September 27, 2021
- Report Date
- December 24, 2021
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- FDS
- UDI-DI
- 04961333237586
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
EVALUATION SUMMARY: THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EG29-I10C-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K190805. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD IMAGE FAILURE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE LEAKAGE OCCURRED IN THE CCD MODULE. IN ADDITION, WE CONFIRMED THAT THE AIR/WATER NOZZLE CLOGGED; HOWEVER, IT IS NOT RELATED TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT, IT WAS EVALUATED TO SUBMIT MDR.
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EG29-I10C-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K190805. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE FILED WITH THE NEW INFORMATION.
FOGGY, MISTY IMAGE. MOISTURE UNDER LED COVER GLASSES. CONNECTION BETWEEN SEGMENT AND IFT LOOSENED, SCREWS MISSING. A/W SOCKET LOOSENED. THIS EVENT OCCURRED AT THE TIME OF DURING INSPECTION. THERE WAS NO REPORT OF PATIENT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1520335 | PENTAX | IMAGINA GASTROSCOPE | FDS | HOYA CORPORATION PENTAX TOKYO OFFICE | EG29-I10C | 04961333237586 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |