FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 12629062 · Received October 13, 2021

Report

Report Number
9610877-2021-01123
Event Type
Malfunction
Date Received
October 13, 2021
Date of Event
September 27, 2021
Report Date
December 24, 2021
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDS
UDI-DI
04961333237586
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

EVALUATION SUMMARY: THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EG29-I10C-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K190805. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD IMAGE FAILURE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE LEAKAGE OCCURRED IN THE CCD MODULE. IN ADDITION, WE CONFIRMED THAT THE AIR/WATER NOZZLE CLOGGED; HOWEVER, IT IS NOT RELATED TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT, IT WAS EVALUATED TO SUBMIT MDR.

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EG29-I10C-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K190805. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE FILED WITH THE NEW INFORMATION.

Description of Event or Problem · 0

FOGGY, MISTY IMAGE. MOISTURE UNDER LED COVER GLASSES. CONNECTION BETWEEN SEGMENT AND IFT LOOSENED, SCREWS MISSING. A/W SOCKET LOOSENED. THIS EVENT OCCURRED AT THE TIME OF DURING INSPECTION. THERE WAS NO REPORT OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1520335 PENTAX IMAGINA GASTROSCOPE FDS HOYA CORPORATION PENTAX TOKYO OFFICE EG29-I10C 04961333237586

Patients

Seq Age Sex Outcome Treatment
1 Unknown