FDA Adverse Event Injury Summary report: N

G7 OSSEOTI 4 HOLE SHELL 64MM H

MDR report key: 12627842 · Received October 13, 2021

Report

Report Number
0001825034-2021-02853
Event Type
Injury
Date Received
October 13, 2021
Date of Event
September 20, 2021
Report Date
December 7, 2021
Manufacturer
ZIMMER BIOMET, INC.
Product Code
LPH
PMA / PMN Number
K140669
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION.

Additional Manufacturer Narrative · 0

MEDICAL RECORDS WERE PROVIDED AND REVIEWED BY A HEALTH CARE PROFESSIONAL. REVIEW OF THE AVAILABLE RECORDS IDENTIFIED THE FOLLOWING: PATIENT UNDERWENT A REVISION SURGERY DUE TO DISLOCATION. ACETABULAR COMPONENT FOUND IN DIFFERENT POSITION FROM TIME OF SURGERY AND MOVES EASILY, WHEN TAPPED, INDICATING DISRUPTED FROM SUBCHONDRAL BONE, EXCHANGED WITH SAME SIZE WITH 3 SCREWS. NEW HEAD , LINER AND SCREWS WERE PLACED. DEVICE HISTORY RECORD WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE APPROXIMATELY 5 DAYS POST IMPLANTATION DUE TO INSTABILITY AND DISLOCATION. DURING THE REVISION THE ACETABULAR SHELL WAS IN A DIFFERENT POSITION FROM PREVIOUS SURGERY. THE SHELL, LINER, BEARING, SLEEVE, AND HEAD WERE EXCHANGED WITHOUT COMPLICATIONS. NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL PRODUCTS: CATALOG#: 110024466 G7 DUAL MOBILITY LINER 50MM H LOT#: 755990. CATALOG#: 110031015 VIVACIT-E DM BEARING 28X50MM LOT#: 64731446. CATALOG#: 650-1068 CER OPTION TYPE 1 TPR SLEVE +6 LOT#: 2900905. CATALOG#: 650-1055 CER BIOLOXD OPTION HD 28MM LOT#: 3050674. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS, DUE TO HOSPITAL POLICY; HOWEVER, AN INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2021-02852, 0001822565-2021-02978.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE DUE TO INSTABILITY AND DISLOCATION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1520259 G7 OSSEOTI 4 HOLE SHELL 64MM H PROSTHESIS, HIP LPH ZIMMER BIOMET, INC. N/A 6459948

Patients

Seq Age Sex Outcome Treatment
1 Male Hospitalization| R