COBAS TAQSCREEN MPX TEST V2
Report
- Report Number
- 2243471-2021-03518
- Event Type
- Malfunction
- Date Received
- October 13, 2021
- Date of Event
- September 14, 2021
- Report Date
- March 4, 2022
- Manufacturer
- ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
- Product Code
- QHO
- UDI-DI
- 00875197004045
- PMA / PMN Number
- BL125459
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TC
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THROUGH THE COURSE OF THE INVESTIGATION, A PRODUCT NON-CONFORMANCE WAS NOT IDENTIFIED WITH THE COMPLAINT KIT LOT. THE SAMPLE WAS PROVIDED FOR FURTHER INVESTIGATION. VIRAL LOAD TESTING ON 2 ALIQUOTS FROM THE SAMPLE WAS PERFORMED BY A THIRD-PARTY EXTERNAL LAB USING THE ABBOTT REALTIME VIRAL LOAD FOR HIV-1 TEST. NO VIRAL TARGET WAS IDENTIFIED. ADDITIONAL TESTING OF THE SAME ALIQUOTS WAS PERFORMED BY ROCHE USING THE COBAS HIV 1/2 QUALITATIVE ASSAY AND BOTH ALIQUOTS GENERATED NON-REACTIVE RESULTS FOR HIV-1 AND HIV-2. ROCHE MEDICAL AND SCIENTIFIC AFFAIRS NOTED THAT HIV-2 INFECTIONS IN DONORS ARE RARE, AND A PREVIOUS STUDY FROM THE AMERICAN RED CROSS REPORTED VERY LOW VIRAL LOADS IN THESE DONORS. THE POTENTIAL AND MOST LIKELY CAUSES FOR THIS PATTERN COULD BE A FALSE POSITIVE GEENIUS RESULT OR A TRUE HIV-2 INFECTION WITH VERY LOW VIRAL LOAD BELOW THE ROCHE TESTS' LIMIT OF DETECTIONS. BASED ON THE TOTALITY OF THE CASE, A PRODUCT PROBLEM WAS NOT IDENTIFIED. CORRECTED DEVICE CODE FROM A090803 FALSE NEGATIVE RESULT TO A090806 TEST RESULT - NON REPRODUCIBLE. (B)(4).
THE INVESTIGATION IS ON-GOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION. (B)(4).
A CUSTOMER IN THE (B)(6) ALLEGED THE GENERATION OF (B)(6) RESULTS WITH THE COBAS TAQSCREEN MPX TEST, V2.0 FOR A PATIENT WHO GENERATED (B)(6) RESULTS. PATIENT WAS BEING SCREENING FOR (B)(6) AS A NEWLY ARRIVED EXPAT. NO DONATION WAS INVOLVED AND NO HARM WAS ALLEGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1522333 | COBAS TAQSCREEN MPX TEST V2 | HUMAN IMMUNODEFICIENCY VIRUS TYPE 1 (HIV-1) GROUP M AND O RNA, HIV-2 RNA, HCV | QHO | ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG | NA | H02619 | 00875197004045 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Male |