FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 12620957 · Received October 13, 2021

Report

Report Number
9610877-2021-50038
Event Type
Malfunction
Date Received
October 13, 2021
Date of Event
September 7, 2021
Report Date
October 13, 2021
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDT
UDI-DI
04961333233007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL ED34-I10T2-US IS AVAILABLE IN THE USA WITH A 510K NUMBER K192245. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE OBJECTIVE PRISM BROKEN. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE OBJECTIVE PRISM.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(OBJECTIVE LENS BROKEN).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1519462 PENTAX VIDEO DUODENOSCOPE FDT HOYA CORPORATION PENTAX TOKYO OFFICE ED34-I10T2 04961333233007

Patients

Seq Age Sex Outcome Treatment
1