FDA Adverse Event Injury Summary report: N

NIPRO SURDIAL DX HEMODIALYSIS SYSTEM

MDR report key: 12609373 · Received October 11, 2021

Report

Report Number
3016250252-2021-00027
Event Type
Injury
Date Received
October 11, 2021
Date of Event
June 24, 2021
Report Date
November 22, 2022
Manufacturer
SHIBUYA (SBY)
Product Code
KDI
PMA / PMN Number
K182940
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITINOAL INFORMATION RECEIVED FROM THE USER: THE NURSE WAS CHECKED BY A PHYSICIAN. SHE RECEIVED MEDICAL TREATMENT/COVERAGE WITH WOUND GAUZE TO SUPPORT THE HEALING ON THE LEG. THE NURSE IS CURRENTLY RECOVERED WITH MINOR MARKS REMAINING ON THE SKIN. SECTION 5 OF THE MANUFACTURER REPORT ATTACHED HAS THE FOLLOWING WARNINGS WHICH ARE ALREADY INCLUDED IN THE MACHINE'S INSTRUCTION MANUAL AND ARE CONTINUALLY REMINDED IN THE INSTRUCTION MANUAL. 6.3. CHECK DURING CLEANING OPERATION. WARNING: THE SURFACES OF EXTERIOR PARTS AND SILICONE TUBES BECOME HOT DURING THE HOT RINSE AND HOT DISINFECTION OPERATIONS. HANDLE WITH CARE NOT TO GET BURN INJURY. THE INSIDE OF THE DIALYSATE CIRCUIT BECOMES VERY HOT DURING THE HOT RINSE AND HOT DISINFECTION OPERATIONS. DO NOT OPERATE WITH UNCOVERED MACHINE. TEMPERATURES OF SOLUTION MAY RISE TO NEARLY 100°C DURING THE HOT RINSE AND HOT DISINFECTION OPERATIONS. DO NOT TOUCH SPILLS WHEN LEAKAGE IS FOUND. YOU MAY GET BURNED.

Description of Event or Problem · 0

WE HAVE BEEN INFORMED ABOUT AN INCIDENT WHERE A NURSE HAD CONTACT WITH HOT WATER WHILE CLEANING THE HEMODIALYSIS MACHINE. THE MACHINE WAS PUT INTO HOT DISINFECTION AND THE NURSE CONTINUED CLEANING / DISINFECTING THE SURFACE OF THE MACHINE. WHILE SHE WAS DOING THIS, SUDDENLY THE SILICONE TUBE FROM THE B-NOZZLE (BICARB NOZZLE) ON THE BSP PANEL (BLOOD PUMP / SYRINGE PUMP PANEL) SIDE "POPPED-OFF" AND HOT WATER WAS SPILLED OVER HER LEG. SHE WAS SUFFERING AND SHOWING SIGNS OF SCALDING. NO ADDITIONAL INFORMATION HAS BEEN RECEIVED AT THIS TIME.

Description of Event or Problem · 1

WE HAVE BEEN INFORMED ABOUT AN INCIDENT WHERE A NURSE HAD CONTACT WITH HOT WATER WHILE CLEANING THE HEMODIALYSIS MACHINE. THE MACHINE WAS PUT INTO HOT DISINFECTION AND THE NURSE CONTINUED CLEANING / DISINFECTING THE SURFACE OF THE MACHINE. WHILE SHE WAS DOING THIS, SUDDENLY THE SILICONE TUBE FROM THE B-NOZZLE (BICARB NOZZLE) ON THE BSP PANEL (BLOOD PUMP / SYRINGE PUMP PANEL) SIDE "POPPED-OFF" AND HOT WATER WAS SPILLED OVER HER LEG. SHE WAS SUFFERING AND SHOWING SIGNS OF SCALDING. NO ADDITIONAL INFORMATION HAS BEEN RECEIVED AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1504290 NIPRO SURDIAL DX HEMODIALYSIS SYSTEM HEMODIALYSIS MACHINE KDI SHIBUYA (SBY)

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other