NIPRO SURDIAL DX HEMODIALYSIS SYSTEM
Report
- Report Number
- 3016250252-2021-00027
- Event Type
- Injury
- Date Received
- October 11, 2021
- Date of Event
- June 24, 2021
- Report Date
- November 22, 2022
- Manufacturer
- SHIBUYA (SBY)
- Product Code
- KDI
- PMA / PMN Number
- K182940
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
ADDITINOAL INFORMATION RECEIVED FROM THE USER: THE NURSE WAS CHECKED BY A PHYSICIAN. SHE RECEIVED MEDICAL TREATMENT/COVERAGE WITH WOUND GAUZE TO SUPPORT THE HEALING ON THE LEG. THE NURSE IS CURRENTLY RECOVERED WITH MINOR MARKS REMAINING ON THE SKIN. SECTION 5 OF THE MANUFACTURER REPORT ATTACHED HAS THE FOLLOWING WARNINGS WHICH ARE ALREADY INCLUDED IN THE MACHINE'S INSTRUCTION MANUAL AND ARE CONTINUALLY REMINDED IN THE INSTRUCTION MANUAL. 6.3. CHECK DURING CLEANING OPERATION. WARNING: THE SURFACES OF EXTERIOR PARTS AND SILICONE TUBES BECOME HOT DURING THE HOT RINSE AND HOT DISINFECTION OPERATIONS. HANDLE WITH CARE NOT TO GET BURN INJURY. THE INSIDE OF THE DIALYSATE CIRCUIT BECOMES VERY HOT DURING THE HOT RINSE AND HOT DISINFECTION OPERATIONS. DO NOT OPERATE WITH UNCOVERED MACHINE. TEMPERATURES OF SOLUTION MAY RISE TO NEARLY 100°C DURING THE HOT RINSE AND HOT DISINFECTION OPERATIONS. DO NOT TOUCH SPILLS WHEN LEAKAGE IS FOUND. YOU MAY GET BURNED.
WE HAVE BEEN INFORMED ABOUT AN INCIDENT WHERE A NURSE HAD CONTACT WITH HOT WATER WHILE CLEANING THE HEMODIALYSIS MACHINE. THE MACHINE WAS PUT INTO HOT DISINFECTION AND THE NURSE CONTINUED CLEANING / DISINFECTING THE SURFACE OF THE MACHINE. WHILE SHE WAS DOING THIS, SUDDENLY THE SILICONE TUBE FROM THE B-NOZZLE (BICARB NOZZLE) ON THE BSP PANEL (BLOOD PUMP / SYRINGE PUMP PANEL) SIDE "POPPED-OFF" AND HOT WATER WAS SPILLED OVER HER LEG. SHE WAS SUFFERING AND SHOWING SIGNS OF SCALDING. NO ADDITIONAL INFORMATION HAS BEEN RECEIVED AT THIS TIME.
WE HAVE BEEN INFORMED ABOUT AN INCIDENT WHERE A NURSE HAD CONTACT WITH HOT WATER WHILE CLEANING THE HEMODIALYSIS MACHINE. THE MACHINE WAS PUT INTO HOT DISINFECTION AND THE NURSE CONTINUED CLEANING / DISINFECTING THE SURFACE OF THE MACHINE. WHILE SHE WAS DOING THIS, SUDDENLY THE SILICONE TUBE FROM THE B-NOZZLE (BICARB NOZZLE) ON THE BSP PANEL (BLOOD PUMP / SYRINGE PUMP PANEL) SIDE "POPPED-OFF" AND HOT WATER WAS SPILLED OVER HER LEG. SHE WAS SUFFERING AND SHOWING SIGNS OF SCALDING. NO ADDITIONAL INFORMATION HAS BEEN RECEIVED AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1504290 | NIPRO SURDIAL DX HEMODIALYSIS SYSTEM | HEMODIALYSIS MACHINE | KDI | SHIBUYA (SBY) |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |