FDA Adverse Event Malfunction Summary report: N

TRANSPAC® IV W/03 ML SQUEEZE FLUSH DEVICE, 60" MACRO ADMIN SET AND PRESSURE TUBI

MDR report key: 12607618 · Received October 11, 2021

Report

Report Number
9617594-2021-00275
Event Type
Malfunction
Date Received
October 11, 2021
Date of Event
September 1, 2021
Report Date
September 20, 2021
Manufacturer
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
Product Code
DRS
UDI-DI
00840619057695
PMA / PMN Number
K061573
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ONE USED LIST #011-46106-74, TRANSPAC® IV W/03 ML SQUEEZE FLUSH DEVICE, 60" MACRO ADMIN SET AND PRESSURE TUBING (152CM); LOT #5142233 WAS RECEIVED FOR EVALUATION. DURING VISUAL INSPECTION, THE PVC TUBING WAS RECEIVED SEPARATED FROM THE WINGED MALE LUER. WHEN THE TUBING POCKET WAS MICROSCOPICALLY EXAMINED, INSUFFICIENT SOLVENT COVERAGE WAS OBSERVED ON THE TUBING. THE PROBABLE CAUSE OF THE DISCONNECTION HAD OCCURRED DUE TO INSUFFICIENT SOLVENT APPLIED ON THE PVC TUBING DURING ASSEMBLY PROCESS. LOT HISTORY REVIEW WAS REVIEWED AND NO NONCONFORMITIES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE REPORTED COMPLAINT.

Additional Manufacturer Narrative · 0

THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION, HOWEVER, IT IS YET TO BE RECEIVED.

Description of Event or Problem · 0

THE EVENT OCCURRED ON AN UNSPECIFIED DATE, AND INVOLVED A TRANSPAC® IV W/03 ML SQUEEZE FLUSH DEVICE, 60" MACRO ADMIN SET AND PRESSURE TUBING (152CM) IN WHICH THE CUSTOMER REPORTED THAT THE TUBING COMES UNDONE AT THE FITTING. THE INCIDENT OCCURRED DURING PRIMING AND THERE WAS NO PATIENT INVOLVEMENT. NO MEDICATION WAS BEING ADMINISTERED; THE DEVICE WAS PRIMED WITH NACL. THE CUSTOMER SWITCHED OUT THE DEVICE AND NO FURTHER ISSUES WERE ENCOUNTERED. THERE WAS NO HARM REPORTED AS CONSEQUENCE OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1501064 TRANSPAC® IV W/03 ML SQUEEZE FLUSH DEVICE, 60" MACRO ADMIN SET AND PRESSURE TUBI TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR DRS ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 5142233 00840619057695

Patients

Seq Age Sex Outcome Treatment
1 Unknown SODIUM CHLORIDE, MFR UNK