GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM
Report
- Report Number
- 2024168-2021-09080
- Event Type
- Injury
- Date Received
- October 11, 2021
- Date of Event
- August 20, 2021
- Report Date
- January 7, 2022
- Manufacturer
- ABBOTT VASCULAR
- Product Code
- MAF
- PMA / PMN Number
- H000001
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
A VISUAL AND DIMENSIONAL INSPECTION WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED FAILURE TO ADVANCE COULD NOT BE TESTED AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY DID NOT INDICATE A LOT SPECIFIC QUALITY ISSUE. THE INVESTIGATION DETERMINED THE REPORTED FAILURE TO ADVANCE APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING.
THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
T WAS REPORTED THAT THE PATIENT PRESENTED WITH AN OPEN PERFORATION IN THE LEFT ANTERIOR DESCENDING (LAD) ARTERY. A 2.8X19MM GRAFTMASTER WAS ATTEMPTED TO BE ADVANCED BUT COULD NOT CROSS THE HEAVILY CALCIFIED ANATOMY TO REACH THE TARGET LESION. THE GRAFTMASTER WAS REMOVED AND ANOTHER GRAFTMASTER WAS USED TO CROSS THE ANATOMY AND SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1502690 | GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM | CORONARY STENT DELIVERY SYSTEM | MAF | ABBOTT VASCULAR | 0091041 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |