CELL-DYN SAPPHIRE ANALYZER
Report
- Report Number
- 2919069-2008-00822
- Event Type
- Malfunction
- Date Received
- December 12, 2008
- Date of Event
- November 18, 2008
- Report Date
- November 18, 2008
- Manufacturer
- ABBOTT DIAGNOSTICS DIVISION/CELLDYN
- Product Code
- GKZ
- PMA / PMN Number
- K051215
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). ON 11/20/2008, THE FIELD SERVICE REPRESENTATIVE (FSR) ATTEMPTED TO REPRODUCE THE PROBLEM WITHOUT SUCCESS. THE FSR REMOVED AND CLEANED THE BAR CODE READER AND THE AREA AROUND IT DUE TO THE PRESENCE OF DUST FILAMENTS IN FRONT OF THE READ WINDOW. THE BAR CODE WAS TESTED AND PASSED INSPECTION. THERE WAS NO REPORT OF AFFECTED PATIENT MANAGEMENT. THERE WAS NO PRODUCT RETURN OR DATA SUBMITTED. THE CELL- DYN SAPPHIRE SYSTEM OPERATOR'S MANUAL, LIST NUMBER 08H10-02, REVISION D, UNDER SECTION 9, SERVICE AND MAINTENANCE PROCEDURES, PAGES 9-50 THROUGH 9-52, DESCRIBES THE PROCEDURE FOR CLEANING THE BAR CODE READER WINDOW. THE MANUAL STATES THAT THE BAR CODE READER WINDOW SHOULD BE CLEANED AS PART OF TROUBLESHOOTING BAR CODE MISREAD. DUST OR DEBRIS CAN "BLIND" THE SENSOR CAUSING THE READER TO MISREAD A LABEL. THE ACTUAL BAR CODE LABEL WAS NOT SUBMITTED FOR INVESTIGATION; THEREFORE, THE BAR CODE CONFIGURATION AND BAR CODE SYMBOLOGY ARE UNKNOWN. THE INFORMATION PROVIDED STATED THAT THE BAR CODE USED IN THE LAB HAD NO CHECKSUM (CHECK DIGIT). CHECK DIGIT IS AN EXTRA NUMERIC CHARACTER IN THE BAR CODE THAT PERMITS THE SCANNING DEVICE TO MATHEMATICALLY DETERMINE WHETHER IT READ THE CODE CORRECTLY. THE CHECK DIGIT HELPS DECREASE BAR CODE LABEL MISREADS. SECTION 2, INSTALLATION PROCEDURES AND SPECIAL REQUIREMENTS, CUSTOMIZING THE SYSTEM FOR BAR CODES, PAGE 2-102, INDICATES THAT THE BAR CODE READER IN THE CELL-DYN SAPPHIRE AUTOLOADER CAN BE CUSTOMIZED TO CALCULATE A CHECK CHARACTER/DIGIT IF ONE IS IN THE BAR CODE SYMBOLOGY. THE OPERATOR CAN SETUP, VERIFY, OR CHANGE THE CHECK CHARACTER/DIGIT SETTINGS FOR THE AUTOLOADER BAR CODE READER. CODE 128 SYMBOLOGY ALWAYS USES A CHECK DIGIT AND THE OPTION TO CUSTOMIZE IS NOT AVAILABLE. A REVIEW OF THE DOMESTIC AND INTERNATIONAL COMPLAINT ANALYSIS TRENDING SYSTEM (CATS) AND TRENDING ANALYSIS SUPPORT SYSTEM'S (TASS) TREND ANALYSIS REPORTS ON THE MTRS DATABASE ON LIST BASE 08H00-01, FOR THE MONTHS OF (B)(6) 2008 DID NOT INDICATE AN ADVERSE TREND FOR THE REPORTED ISSUE. BASED ON THIS INVESTIGATION, NO PRODUCT DEFICIENCY/MALFUNCTION WAS IDENTIFIED FOR THE CELL-DYN SAPPHIRE SYSTEM, LIST NUMBER 08H00-01, FOR THE REPORTED ISSUE. THIS IS THE FINAL REPORT.
THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
THE ACCOUNT STATED THAT THE CELL-DYN SAPPHIRE ANALYZER INCORRECTLY READ THE BARCODE ON A SAMPLE THE SAMPLE HAD A BARCODE ID OF 935 AND THE SAPPHIRE READ THE SAMPLE AS 955. UPON READING THE SAMPLE AGAIN IT WAS CORRECTLY READ AS 935. THE FIELD SERVICE REPRESENTATIVE (FSR) WAS DISPATCHED AND CLEANED THE BARCODE READER, AND THE AREA AROUND IT AS THERE WAS THE PRESENCE OF DUST IN FRONT OF THE READING WINDOW. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CELL-DYN SAPPHIRE ANALYZER | AUTOMATED HEMATOLOGY ANALYZER | GKZ | ABBOTT DIAGNOSTICS DIVISION/CELLDYN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |