FDA Adverse Event Injury Summary report: N

ONE TOUCH BASIC

MDR report key: 126034 · Received October 10, 1997

Report

Report Number
2939301-1997-00071
Event Type
Injury
Date Received
October 10, 1997
Date of Event
September 11, 1997
Report Date
September 12, 1997
Manufacturer
LIFESCAN INC.
Product Code
CFR
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

CO IS UNABLE TO COMPLETE ITS INVESTIGATION OF THE MDR INCIDENT LISTED ABOVE DUE TO THE FACT THAT THE METER WAS NOT RETURNED TO LIFESCAN. THEREFORE, CO HAS NOT BEEN ABLE TO EVALUATE THE METER TO DETERMINE WHETHER OR NOT A DEVICE MALFUNCTION MAY HAVE CONTRIBUTED TO THIS EVENT. BATCH RECORD REVIEW INDICATES NO NON-CONFORMANCES IN THE MFG PROCESS. AT THIS POINT, CO CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

THE PT WAS FEELING ILL WITH NAUSEA, WEAKNESS AND DEEP SLEEP ON 9/11. A BLOOD GLUCOSE TEST PERFORMED ON HER ONE TOUCH BASIC METER AT 5:30/P WAS 45 MG/DL. SHE WAS ADVISED TO CONSUME 8 OZ. ORANGE JUCE. BECAUSE THERE WAS NO IMPROVEMENT IN HER SYMPTOMS, SHE WAS TRANSPORTED TO THE HOSP WHERE A BLOOD GLUCOSE READING OF 654 MG/DL WAS OBTAINED (METHOD UNK). THE PT WAS ADMITTED FOR ELEVATED GLUCOSE AND A BLADDER INFECTION AND TREATED WITH IV FLUIDS AND INSULIN INJECTIONS (OTHER TREATMENT UNK). THE METER IS NOT CLEANED OR MAINTAINED ACCORDING TO MFR'S RECOMMENDATIONS. ALCOHOL IS USED TO CLEAN THE METER OPTICS. THE PRODUCT'S INSTRUCTIONS FOR USE STATE THAT USE OF ALCOHOL AS A CLEANING AGENT CAN DAMAGE THE METER OPTICS. QUALITY CONTROL TESTS DESIGNED TO DETECT POTENTIALLY INACCURATE RESULTS ARE NOT PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH BASIC BLOOD GLUCOSE METER CFR LIFESCAN INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 74 YR Hospitalization| R