FDA Adverse Event Malfunction Summary report: N

DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 12599448 · Received October 8, 2021

Report

Report Number
3004753838-2021-183279
Event Type
Malfunction
Date Received
October 8, 2021
Date of Event
September 10, 2021
Report Date
October 29, 2021
Manufacturer
DEXCOM, INC.
Product Code
QBJ
PMA / PMN Number
DEN170088
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(4).

Additional Manufacturer Narrative · 0

(B)(4). 3004753838 -2021 -183279 -FOLLOW-UP 001 THE SUPPLEMENTAL WAS REPORTED IN ERROR. PLEASE DISREGARD SUPPLEMENTAL REPORTING OF THIS EVENT AS THIS EVENT HAS BEEN REPORTED IN THE INITIAL SUBMISSION. H2 IF FOLLOW-UP, WHAT TYPE-CORRECTION.

Additional Manufacturer Narrative · 0

(B)(4) 3004753838 -2021 -183279 -FOLLOW-UP 001 THE SUPPLEMENTAL WAS REPORTED IN ERROR. PLEASE DISREGARD SUPPLEMENTAL REPORTING OF THIS EVENT AS THIS EVENT HAS BEEN REPORTED IN THE INITIAL SUBMISSION.

Description of Event or Problem · 0

IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE MEASUREMENT WAS PERFORMED AND PASSED. DOWNLOAD/PAIRING WAS PERFORMED AND FAILED. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND TRANSMITTER FAILED ERROR WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE MEASUREMENT WAS PERFORMED AND PASSED. DOWNLOAD/PAIRING WAS PERFORMED AND FAILED. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND TRANSMITTER FAILED ERROR WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE MEASUREMENT WAS PERFORMED AND PASSED. DOWNLOAD/PAIRING WAS PERFORMED AND FAILED. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND TRANSMITTER FAILED ERROR WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT TRANSMITTER FAILED ERROR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE MEASUREMENT WAS PERFORMED AND PASSED. DOWNLOAD/PAIRING WAS PERFORMED AND FAILED. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND TRANSMITTER FAILED ERROR WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.