670G INSULIN PUMP MMT-1780KL
Report
- Report Number
- 2032227-2021-202865
- Event Type
- Injury
- Date Received
- October 8, 2021
- Date of Event
- September 15, 2021
- Report Date
- February 28, 2022
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000365882
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
RETAINER RING = BLACK. ON (B)(6) 2021 THE CUSTOMER ALLEGED PUMP UNDER DELIVERING AND WAS HOSPITALIZED FOR HIGH BGS. UNIT PASSED THE FUNCTIONAL TEST, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST AND THE DAT TEST AT 0.0875 INCHES. NO UNDER DELIVERY ANOMALY OR OVER DELIVERY ANOMALY NOTED. THERE WAS A RELATED SVN WITH AN EVENT DATE OF (B)(6) 2021. THERE WAS OVER 400 BOLUSES DELIVERED BETWEEN (B)(6) 2021 TO (B)(6) 2021. REVIEWED THE FIRST 10 BOLUSES ON EACH DAY. (B)(6) 2021 00:04:21.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.075. BOLUS AMOUNT DELIVERED = 0.075. (B)(6) 2021 00:09:18.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.05. BOLUS AMOUNT DELIVERED = 0.05. (B)(6) 2021 00:14:19.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.075. BOLUS AMOUNT DELIVERED = 0.075. 09/14/2021 01:24:17.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.075. BOLUS AMOUNT DELIVERED = 0.075. (B)(6) 2021 01:29:15.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.075. BOLUS AMOUNT DELIVERED = 0.075, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). (B)(6) 2021 01:34:13.000 NORMAL BOLUS DELIVERED. NORMAL BOLUS AMOUNT PROGRAMMED = 0.025. BOLUS AMOUNT DELIVERED = 0.025. (B)(6) 2021 01:39:12.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.025. BOLUS AMOUNT DELIVERED = 0.025. (B)(6) 2021 01:44:27.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.075. BOLUS AMOUNT DELIVERED = 0.075. (B)(6) 2021 01:49:11.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.075. BOLUS AMOUNT DELIVERED = 0.075. (B)(6) 2021 01:54:25.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.075. BOLUS AMOUNT DELIVERED = 0.075. (B)(6) 2021 00:00:00.000 DAILY TOTALS. EVENT TYPE = 63. SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). HISTORY RECORD LENGTH = 114. SYSTEM TIME = (B)(6) 2021 00:00:00.000. DAILY TOTAL COLLECTION START TIME = 09/14/2021 00:00:00.000. DURATION = 2160. NUMBER OF BG = 18. AVERAGE BG = 310. LOW BG = 157. HIGH BG = 397. BG MANUAL AND RECEIVED STANDARD DEVIATION = 64. DAILY TOTAL OF ALL INSULIN DELIVERED = 48.45. DAILY TOTAL OF BASAL INSULIN DELIVERED = 24.25. PERCENTAGE OF DAILY TOTAL DELIVERED AS BASAL INSULIN = 50. DAILY TOTAL OF BOLUS INSULIN DELIVERED = 24.2. (B)(6) 2021 00:04:14.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.05. BOLUS AMOUNT DELIVERED = 0.05. (B)(6) 2021 00:09:11.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.025. BOLUS AMOUNT DELIVERED = 0.025. (B)(6) 2021 00:16:02.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.05. BOLUS AMOUNT DELIVERED = 0.05. (B)(6) 2021 00:21:02.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.05. BOLUS AMOUNT DELIVERED = 0.05. (B)(6) 2021 00:26:02.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.05. BOLUS AMOUNT DELIVERED = 0.05. (B)(6) 2021 00:31:02.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.025. BOLUS AMOUNT DELIVERED = 0.025. (B)(6) 2021 00:36:02.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.05. BOLUS AMOUNT DELIVERED = 0.05. (B)(6) 2021 00:46:02.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.05. BOLUS AMOUNT DELIVERED = 0.05. (B)(6) 2021 00:51:02.000 NORMAL BOLUS DELIVERED. NORMAL BOLUS AMOUNT PROGRAMMED = 0.025. BOLUS AMOUNT DELIVERED = 0.025. (B)(6) 2021 00:56:02.000 NORMAL BOLUS DELIVERED, SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE). NORMAL BOLUS AMOUNT PROGRAMMED = 0.05. BOLUS AMOUNT DELIVERED = 0.05. (B)(6) 2021 00:00:00.000 DAILY TOTALS. EVENT TYPE = 63. SOURCE ID = PUMP (NGP IS IN CLOSED LOOP ACTIVE STATE) HISTORY RECORD LENGTH = 114. SYSTEM TIME = (B)(6) 2021 00:00:00.000. DAILY TOTAL COLLECTION START TIME = 09/15/2021 00:00:00.000. DURATION = 1440. NUMBER OF BG = 2. AVERAGE BG = 346. LOW BG = 298. HIGH BG = 393. BG MANUAL AND RECEIVED STANDARD DEVIATION = 47. DAILY TOTAL OF ALL INSULIN DELIVERED = 28.525. DAILY TOTAL OF BASAL INSULIN DELIVERED = 21.025. PERCENTAGE OF DAILY TOTAL DELIVERED AS BASAL INSULIN = 74. DAILY TOTAL OF BOLUS INSULIN DELIVERED = 7.5. NO BOLUS ANOMALY NOTED. HISTORY DOWNLOAD WAS SUCCESSFUL USING THUS AND CARELINK UPLOAD WAS SUCCESSFUL. UNIT HAD SCRATCHED CASE AND A PILLOWING KEYPAD OVERLAY. THE TEST P-CAP AND RESERVOIR DOES LOCK IN PLACE IN THE RESERVOIR COMPARTMENT. IN SUMMARY, CUSTOMER ALLEGATION FOR PUMP UNDER DELIVERING NOT CONFIRMED DURING FULL FUNCTIONAL TESTING. UNABLE TO VERIFY CUSTOMER ALLEGED FOR HIGH BGS. (B)(4). MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
(B)(4). A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.
THE CUSTOMER REPORTED VIA PHONE CALL THAT CUSTOMER WAS HOSPITALIZED AFTER EMERGENCY MEDICAL SERVICES WERE DISPATCHED DUE TO HIGH BLOOD GLUCOSE ON (B)(6) 2021. BLOOD GLUCOSE AT THE TIME OF EVENT WAS UNKNOWN. CURRENT BLOOD GLUCOSE WAS 350 MG/DL. THE CUSTOMER DID EXPERIENCE SYMPTOMS SUCH AS NAUSEA AND VOMITING A RESULT OF HIGH BLOOD GLUCOSE. CUSTOMER HAD BEEN USING INSULIN PUMP SYSTEM WITHIN 48 HOURS OF REPORTED HIGH BLOOD GLUCOSE EVENT. CUSTOMER DID NOT USE AUTO MODE FEATURE AT THE TIME OF HIGH BLOOD GLUCOSE EVENT. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1494778 | 670G INSULIN PUMP MMT-1780KL | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1780KL | HG5LRFR | 000000763000365882 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Female | Other | FRN-UNK-RSVR, UNOMED INF SET| FRN-UNK-RSVR, UNOMED INF SET |