FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19

MDR report key: 12589511 · Received October 7, 2021

Report

Report Number
1221359-2021-03068
Event Type
Malfunction
Date Received
October 7, 2021
Date of Event
September 14, 2021
Report Date
November 16, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
04571226475027
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M161624 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING AND QUALITY CONTROL RELEASE TESTING WERE REVIEWED FOR TEST BASE LOT M161624 AND THEY MET RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M161624 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE, HOWEVER, IT COULD POSSIBLY BE RELATED TO SAMPLE INTERFERENCE.

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 PERFORMED ON (B)(6) 2021 ON A BIPRO SPONGE SWAB TYPE S. REPEAT TESTING WAS NOT PERFORMED. CONFIRMATION TESTING ON NASOPHARYNGEAL SWABS WITH SHIMADZU AUTOAMP ONE-STEP MULTIPLEX RT-PCR HYDROLYSIS PROBE GENERATED NEGATIVE RESULTS (CT VALUES NOT PROVIDED). THE CUSTOMER STATED THE PATIENT WAS ASYMPTOMATIC. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1488843 ID NOW COVID-19 MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. M161624 04571226475027

Patients

Seq Age Sex Outcome Treatment
1 93 YR Male