FDA Adverse Event Malfunction Summary report: N

CEPHEID XPERT XPRESS SARS-COV-2/FLU/RSV

MDR report key: 12586626 · Received October 6, 2021

Report

Report Number
3004530258-2021-00407
Event Type
Malfunction
Date Received
October 6, 2021
Date of Event
September 16, 2021
Report Date
October 6, 2021
Manufacturer
CEPHEID
Product Code
QJR
PMA / PMN Number
EUA200453
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

US CUSTOMER CONTACTED CEPHEID TO DISCUSS PATIENT RESULTS FOR XPERT XPRESS SARS-COV-2/FLU/RSV ON THE SITE'S GENEXPERT INSTRUMENT. CUSTOMER COLLECTED SAMPLE 1 FROM PATIENT ON (B)(6) 2021, WHICH WAS TESTED ON XPERT XPRESS SARS-COV-2/FLU/RSV AND RESULTED AS SARS-COV-2 NEGATIVE, FLU A POSITIVE, FLU B NEGATIVE, RSV NEGATIVE (REPORTED TO THE PHYSICIAN). SAMPLE 1 REPEAT 1 OCCURED ON (B)(6) 2021 AND TESTED ON XPERT XPRESS SARS-COV-2/FLU/RSV, WHICH RESULTED IN SARS-COV-2 NEGATIVE, FLU A NEGATIVE, FLU B NEGATIVE, RSV NEGATIVE (REPORTED TO THE PHYSICIAN). THE ROOT CAUSE WAS DETERMINED TO BE CONTAMINATION AND/OR CARRYOVER. THERE IS NO EVIDENCE OF PRODUCT MALFUNCTION AND THERE WAS NO REPORTED HARM. FALSE POSITIVE RESULTS FOR SARS-COV-2 COULD LEAD TO INCORRECT MANAGEMENT OF SYMPTOMATIC PATIENTS, INCLUDING UNNECESSARY ISOLATION OR QUARANTINE, MISALLOCATION OF RESOURCES, DELAYED DIAGNOSIS AND TREATMENT FOR OTHER INFECTIONS OR HEALTH CONDITIONS, OR UNNECESSARY ANTIVIRAL TREATMENT.  CEPHEID'S PATIENT SAFETY BOARD CONSULT CONFIRMED WITH THE CUSTOMER THAT THERE WAS NO ASSOCIATED PATIENT HARM OR CHANGE TO PATIENT TREATMENT. RESULTS ARE FOR THE IDENTIFICATION OF SARS-COV-2 RNA.  AS PER SECTION 3 OF XPERT XPRESS SARS-COV-2/FLU/RSV EUA IFU; POSITIVE RESULTS ARE INDICATIVE OF THE PRESENCE OF SARS-COV-2; CLINICAL CORRELATION WITH PATIENT HISTORY AND OTHER DIAGNOSTIC INFORMATION IS NECESSARY TO DETERMINE PATIENT INFECTION STATUS. NOTE DEVICE EVALUATED BY MANUFACTURER - ANSWER OF NO IS DUE TO THE SINGLE USE OF THE XPERT XPRESS SARS-COV-2/FLU/RSV TEST AND THE UNAVAILABILITY OF THAT PRODUCT LOT TO BE RETURNED TO CEPHEID.

Description of Event or Problem · 0

US CUSTOMER CONTACTED CEPHEID TO DISCUSS PATIENT RESULTS FOR XPERT XPRESS SARS-COV-2/FLU/RSV ON THE SITE'S GENEXPERT INSTRUMENT. CUSTOMER COLLECTED SAMPLE 1 FROM PATIENT ON (B)(6) 2021, WHICH WAS TESTED ON XPERT XPRESS SARS-COV-2/FLU/RSV AND RESULTED AS SARS-COV-2 NEGATIVE, FLU A POSITIVE, FLU B NEGATIVE, RSV NEGATIVE (REPORTED TO THE PHYSICIAN). SAMPLE 1 REPEAT 1 OCCURED ON (B)(6) 2021 AND TESTED ON XPERT XPRESS SARS-COV-2/FLU/RSV, WHICH RESULTED IN SARS-COV-2 NEGATIVE, FLU A NEGATIVE, FLU B NEGATIVE, RSV NEGATIVE (REPORTED TO THE PHYSICIAN). REVIEW OF DATA PROVIDED BY THE CUSTOMER SHOWED NO EVIDENCE OF PRODUCT MALFUNCTION AND THERE WAS NO KNOWN HARM TO PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1483590 CEPHEID XPERT XPRESS SARS-COV-2/FLU/RSV COVID-19 RT-PCR TEST QJR CEPHEID 1000282838

Patients

Seq Age Sex Outcome Treatment