FDA Adverse Event Malfunction Summary report: N

OPTILENE 5/0 (1) 75CM 2XDRC13 CV2 RCP

MDR report key: 12582351 · Received October 6, 2021

Report

Report Number
3003639970-2021-00505
Event Type
Malfunction
Date Received
October 6, 2021
Report Date
December 1, 2021
Manufacturer
B. BRAUN SURGICAL, S.A.U.
Product Code
GAW
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE-BATCH. WE MANUFACTURED AND DISTRIBUTED IN THE MARKET (B)(4) UNITS OF THIS CODE-BATCH. THERE ARE NO UNITS IN OUR STOCK. WE HAVE RECEIVED 77 CLOSED SAMPLES TO ANALYZE THIS COMPLAINT. WE HAVE TESTED THE NEEDLE ATTACHMENT STRENGTH OF THE SAMPLES RECEIVED AND THE RESULTS DO NOT FULFIL THE REQUIREMENTS OF THE EUROPEAN PHARMACOPOEIA (EP). REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFILLED USP/EP AND B.BRAUN SURGICAL REQUIREMENTS. FINAL CONCLUSION: TAKING INTO ACCOUNT THAT THE RESULTS OF SAMPLES RECEIVED DO NOT FULFIL THE SPECIFICATIONS OF THE EUROPEAN PHARMACOPOEIA/B. BRAUN SURGICAL SPECIFICATIONS, WE CONCLUDE THAT THE COMPLAINT IS CONFIRMED. ACTIONS ON PRODUCT DISTRIBUTED OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. CORRECTIVE/PREVENTIVE ACTIONS: ACCORDING TO OUR INTERNAL PROCEDURES, THERE IS NO NEED TO ESTABLISH CORRECTIVE OR PREVENTIVE ACTIONS. NEVERTHELESS, THIS COMPLAINT IS RECORDED FOR TRENDING ANALYSIS TO ASSESS IF ACTIONS ARE NEEDED IN THE FUTURE.

Additional Manufacturer Narrative · 1

REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. UNDER PMA/510(K) NUMBER: K133890. IF ADDITIONAL INFORMATION BECOMES AVAILABLE A FOLLOW UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED AN ISSUE WITH OPTILENE SUTURE. THE CLIENT REPORTED THAT THE NEEDLE DETACHED FROM THE THREAD 4 CONSECUTIVE TIMES DURING AORTIC AND VENOUS BYPASS SURGERY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1479967 OPTILENE 5/0 (1) 75CM 2XDRC13 CV2 RCP CARDIAC SUTURE GAW B. BRAUN SURGICAL, S.A.U. C3097141 121225

Patients

Seq Age Sex Outcome Treatment
1 71 YR Male