FDA Adverse Event Injury Summary report: N

IMP,TSV,4.1MM,SBM,11.5

MDR report key: 12579478 · Received October 5, 2021

Report

Report Number
0002023141-2021-02767
Event Type
Injury
Date Received
October 5, 2021
Date of Event
September 21, 2021
Report Date
February 25, 2022
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019232
PMA / PMN Number
K072589
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CS
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SIMILAR COMPLAINTS FOR IMPLANT INFECTION HAVE BEEN PREVIOUSLY INVESTIGATED. VISUAL AND DIMENSIONAL EVALUATIONS OF THE PREVIOUSLY RETURNED PRODUCT HAVE NOT IDENTIFIED OR SUGGESTED MANUFACTURING NON-CONFORMANCES. WHILE NON-CONFORMANCES WERE IDENTIFIED FOR SOME LOTS DURING MANUFACTURING RECORD REVIEWS, THE DOCUMENTED DISPOSITION ACTIONS FOR EACH DID NOT SUGGEST THE LIKELY RELEASE OF NON-CONFORMING PRODUCT. ZIMMER BIOMET QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT. ADDITIONALLY, ALL DEVICE HISTORY RECORD REVIEWS VERIFIED THAT EACH IMPLANT WAS STERILIZED PER PROCEDURE FOR EVERY DEVICE. ALL COMPLAINT DATA USED FOR THE SUMMARY INVESTIGATION WAS FOUND TO BE CONFORMING AND DID NOT MEET CAPA/HHE/D/IE ESCALATION. THEREFORE, THERE WERE NO COMPLAINTS WHICH CONFIRMED A MANUFACTURING OR DESIGN RELATED ISSUE THAT DID OR COULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1233258). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WERE NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBERS (1233258) FOR SIMILAR EVENTS AND 1 OTHER COMPLAINT WAS IDENTIFIED. AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATION, NO MALFUNCTION OCCURRED UPON INVESTIGATION. THE REPORTED EVENT REMAINS NON-VERIFIABLE AS IT IS A MEDICAL CONDITION.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). PATIENT WEIGHT UNKNOWN / NOT PROVIDED. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR INFECTION EVENTS IDENTIFYING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO THE WIDE RANGE OF EXTERNAL (NON-DESIGN/ NON-MANUFACTURING RELATED) FACTORS POTENTIALLY IMPACTING THE STERILITY OF THE IMPLANT AND ITS ENVIRONMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE INFECTION EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DR. INDICATED INFECTION. THE IMPLANT WAS PLACED ON (B)(6) 2020, 8 MONTHS LATER THE PATIENT PRESENT WITH INFECTION IN THE IMPLANT AREA WHICH CAUSE THE LOST. SYMPTOMS AS A RESULT OF THE EVENT: PAIN, INFLAMMATION. TOOTH SITE # 26.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1473309 IMP,TSV,4.1MM,SBM,11.5 DENTAL IMPLANT DZE ZIMMER DENTAL TSV4B11 1233258 00889024019232

Patients

Seq Age Sex Outcome Treatment
1 67 YR Female Required Intervention