FDA Adverse Event Malfunction Summary report: N

PERKINELMER NEW CORONAVIRUS NUCLEIC ACID DETECTION KIT

MDR report key: 12576193 · Received October 4, 2021

Report

Report Number
MW5104369
Event Type
Malfunction
Date Received
October 4, 2021
Report Date
September 30, 2021
Manufacturer
PERKINELMER, INC.
Product Code
QJR
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

PERKINELMER NEW CORONAVIRUS NUCLEIC ACID DETECTION KIT USE FOR COVID-19 UNDER EMERGENCY USE AUTHORIZATION (EUA): WE HAVE IDENTIFIED AN ISSUE WITH THE BD UNIVERSAL VIRAL TRANSPORT MEDIUM FOR VIRUSES, CHLAMYDIAE, MYCOPLASMAS, UREAPLASMAS AND THE PERKINELMER NEW CORONAVIRUS NUCLEIC ACID DETECTION KIT. WE WERE ALERTED TO THIS ISSUE WHEN A SUBMITTER TRANSITIONED TO USING THE BD UTM AND WE IMMEDIATELY SAW AN INCREASE IN THE NUMBER OF INVALID RESULTS. WE HAVE INVESTIGATED THOROUGHLY AND HAVE COME TO THE CONCLUSION THAT IT'S THIS PARTICULAR MEDIA THAT IS CAUSING THE ISSUE. THE RESULTS ARE REPEATABLE FOLLOWING RE-EXTRACTION. THE 2 LOTS WE HAVE TESTED ARE: BD/N101148 AND BD/N100701. FDA SAFETY REPORT ID# (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1464334 PERKINELMER NEW CORONAVIRUS NUCLEIC ACID DETECTION KIT REAGENTS, 2019-NOVEL CORONAVIRUS NUCLEIC QJR PERKINELMER, INC.
1464336 THE BD UNIVERSAL VIRAL TRANSPORT MEDIUM FOR VIRUSES, CHLAMYDIAE, MYCOP CULTURE MEDIA, NON-PROPAGATING TRANSPORT JSM COPAN DIAGNOSTICS, INC. BD/ N101148
1464337 THE BD UNIVERSAL VIRAL TRANSPORT MEDIUM FOR VIRUSES, CHLAMYDIAE, MYCOP CULTURE MEDIA, NON-PROPAGATING TRANSPORT JSM COPAN DIAGNOSTICS, INC. BD/ N100701

Patients

Seq Age Sex Outcome Treatment
1 Other