CHLORAPREP ONE STEP HI-LITE ORANGE
Report
- Report Number
- 3004932373-2021-00454
- Event Type
- Malfunction
- Date Received
- October 4, 2021
- Date of Event
- September 7, 2021
- Report Date
- November 2, 2021
- Product Code
- KXF
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PHOTOS WERE AVAILABLE FOR EVALUATION. VISUAL EXAMINATION OF THE PHOTOS SHOWS FOREIGN MATERIAL ON THE SPONGE OF THE APPLICATORS PICTURED. THIS VERIFIED THE REPORTED ISSUE. THE MOST PROBABLE ROOT CAUSE IS INADEQUATE GOWNING BY ASSOCIATE(S) AND/OR PREVENTIVE MEASURES DURING THE PACKAGING OF THE PRODUCT SUCH CLEANING ACTIVITIES. THE PRODUCTION RECORD REVIEW WAS COMPLETED FOR BATCH/LOT 0170675 AND NO NON-CONFORMITIES, FAILURES, DEVIATIONS, OR REWORK ACTIVITIES OCCURRED DURING THE MANUFACTURING OF THIS LOT SIMILAR TO THE REPORTED ISSUE OR THAT COULD HAVE CONTRIBUTED TO THE REPORTED FAILURE MODE. RECORDS REVIEWED INDICATE THAT THE LOT PASSED ALL THE IN-PROCESS INSPECTIONS. NO FURTHER ACTIONS ARE REQUIRED. THIS FAILURE WILL CONTINUE TO BE TRACKED AND TRENDED.
THEY REPORT THAT DURING THE MANUFACTURING PROCESS THEY FIND 6 PIECES THAT PRESENT FOREIGN PARTICLES ON THE SPONGES. ___ FOREIGN PARTICLES: PR# (B)(4). SPOTS/DOTS: PR# (B)(4). *** ADDITIONAL INFORMATION RECEIVED ON 10/07/2021: OF THE 6 PIECES: 5 PIECES WITH FOREIGN PARTICLES AND 1 PIECE WITH SPOTS. ****
(B)(4) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
THEY REPORT THAT DURING THE MANUFACTURING PROCESS THEY FIND 6 PIECES THAT PRESENT FOREIGN PARTICLES ON THE SPONGES. FOREIGN PARTICLES: (B)(4). SPOTS/DOTS: (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1470545 | CHLORAPREP ONE STEP HI-LITE ORANGE | 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL | KXF | 0170675 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |