FDA Adverse Event
Malfunction
Summary report: N
CRYOSTORE CS500N
MDR report key: 12561773
·
Received September 30, 2021
Report
- Report Number
- MW5104308
- Event Type
- Malfunction
- Date Received
- September 30, 2021
- Date of Event
- September 27, 2021
- Report Date
- September 28, 2021
- Manufacturer
- ORIGEN BIOMEDICAL, INC
- Product Code
- KSR
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
ORIGEN CRYOSTORE CS500N BAG WAS USED TO CRYOPRESERVE HPC, APHERESIS AUTOLOGOUS PRODUCT FOR A PATIENT. 100ML WAS CRYOPRESERVED IN THE BAG AND KEPT IN LIQUID NITROGEN VAPOR PHASE STORAGE. NO PROBLEMS WERE FOUND DURING CRYOPRESERVATION. THE BAG WAS THAWED AT THE PATIENT'S BEDSIDE. ONCE COMPLETELY THAWED, THE BAG WAS TURNED UPSIDE DOWN FOR INFUSION AND A LEAK WAS NOTICED FROM THE BAG. NO LEAK WAS SEEN DURING THAWING. A HOLE/BREAK BENEATH THE PORT WAS FOUND AS THE SOURCE FOR THE LEAK. THE PRODUCT WAS INFUSED WITH A SYRINGE TO PREVENT FURTHER LOSS. FDA SAFETY REPORT ID # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1449527 | CRYOSTORE CS500N | CONTAINER, EMPTY, FOR COLLECTION & PROCESSING OF BLOOD & BLOOD COMPONENTS | KSR | ORIGEN BIOMEDICAL, INC | CS500N | W21779 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |