FDA Adverse Event Malfunction Summary report: N

CRYOSTORE CS500N

MDR report key: 12561773 · Received September 30, 2021

Report

Report Number
MW5104308
Event Type
Malfunction
Date Received
September 30, 2021
Date of Event
September 27, 2021
Report Date
September 28, 2021
Manufacturer
ORIGEN BIOMEDICAL, INC
Product Code
KSR
Product Problem
Yes
Report Source
Voluntary report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

ORIGEN CRYOSTORE CS500N BAG WAS USED TO CRYOPRESERVE HPC, APHERESIS AUTOLOGOUS PRODUCT FOR A PATIENT. 100ML WAS CRYOPRESERVED IN THE BAG AND KEPT IN LIQUID NITROGEN VAPOR PHASE STORAGE. NO PROBLEMS WERE FOUND DURING CRYOPRESERVATION. THE BAG WAS THAWED AT THE PATIENT'S BEDSIDE. ONCE COMPLETELY THAWED, THE BAG WAS TURNED UPSIDE DOWN FOR INFUSION AND A LEAK WAS NOTICED FROM THE BAG. NO LEAK WAS SEEN DURING THAWING. A HOLE/BREAK BENEATH THE PORT WAS FOUND AS THE SOURCE FOR THE LEAK. THE PRODUCT WAS INFUSED WITH A SYRINGE TO PREVENT FURTHER LOSS. FDA SAFETY REPORT ID # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1449527 CRYOSTORE CS500N CONTAINER, EMPTY, FOR COLLECTION & PROCESSING OF BLOOD & BLOOD COMPONENTS KSR ORIGEN BIOMEDICAL, INC CS500N W21779

Patients

Seq Age Sex Outcome Treatment
1 65 YR