MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL
Report
- Report Number
- 2032227-2021-200105
- Event Type
- Malfunction
- Date Received
- September 30, 2021
- Date of Event
- September 22, 2021
- Report Date
- March 30, 2022
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000439859
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
RETAINER RING = BLACK. CUSTOMER RETURNED PUMP FOR AN ALLEGED COSMETIC DAMAGE LOCATED AT THE BATTERY COMPARTMENT AND BLANK DISPLAY FOUND ON (B)(6) 2021. PUMP WAS RECEIVED WITH A PARTIALLY BROKEN BATTERY TUBE THREADS. THE PUMP POWERED UP PROPERLY AFTER BATTERY INSTALLATION. THE PUMP PASSED THE DISPLACEMENT TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT AND SELF TEST. THE PUMP WAS MONITORED FOR SEVERAL DAYS AND NO BLANK DISPLAY NOTED. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUMP. POWER MANAGEMENT GRAPH WAS SUCCESSFULLY GENERATED. THE POWER MANAGEMENT TOOL CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. NO ALARMS/ALERTS WERE NOTED DURING THE TESTING. NO ALARMS/ALERTS WERE FOUND IN THE HISTORY FILES OR TRACES FOR THE EVENT DATE. HOWEVER, INSERT BATTERY ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON: (B)(6) 2021 05:04:12.000. ON (B)(6) 2021 08:19:23.000 AND ON (B)(6) 2021 08:29:00.000. LOW BATTERY ALERT WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON (B)(6) 2021 23:44:00.000. ON (B)(6) 2021 19:31:00.000. REPLACE BATTERY ALERT WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON (B)(6) 2021 05:56:00.000. ON (B)(6) 2021 05:02:00.000 AND ON (B)(6) 2021 05:12:00.000. REPLACE BATTERY NOW ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON (B)(6) 2021 06:27:00.000. ON (B)(6) 2021 05:33:00.000 AND ON (B)(6) 2021 05:43:00.000. POWER LOSS ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON (B)(6) 2021 06:38:43.000 AND ON (B)(6) 2021 06:38:51.000. ON (B)(6) 2021 05:44:34.000 TO ON (B)(6) 2021 08:41:02.000. AND FAILED BATT/BATTERY FAILED ALARM WAS RECORDED AND FOUND IN THE FORMATTED HISTORY FILE ON (B)(6) 2021 05:05:30.000. UPON CHECKING ON THE POWER DATA, LOW BATTERY ALERT, REPLACE BATTERY ALERT, REPLACE BATTERY NOW ALARM AND POWER LOSS ALARM WAS EXPECTED SINCE THE BATTERY HAS LOW/NO POWER. THE CUSTOMER MAY HAVE USED A LOW/DEPLETED BATTERY. THE PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND IT WAS VERIFIED THAT THE BATTERY TUBE POWER CONNECTOR IS PROPERLY CONNECTED TO J7/PCB1. NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE ON THE ELECTRONIC ASSEMBLY, FORCE SENSOR, MOTOR AND VIBRATOR ASSEMBLY NOTED. HOWEVER, CORROSION WAS FOUND ON THE BATTERY TUBE ASSEMBLY NOTED. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A CRACKED KEYPAD OVERLAY, A PILLOWING KEYPAD OVERLAY AND A SCRATCHED CASE. COSMETIC DAMAGE WAS CONFIRMED AT THE BATTERY COMPARTMENT. BLANK DISPLAY WAS NOT CONFIRMED. HOWEVER, CORROSION WAS FOUND ON THE BATTERY TUBE ASSEMBLY NOTED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD BLANK DISPLAY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1448878 | MINIMED 770G US SYSTEM BLE CONNECT 3.0 MG/DL | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1880 | HG4Z0QDZZ | 000000763000439859 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Male |