FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19

MDR report key: 12552358 · Received September 30, 2021

Report

Report Number
1221359-2021-02927
Event Type
Malfunction
Date Received
September 30, 2021
Date of Event
August 18, 2021
Report Date
December 8, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCABOROUGH
Product Code
QJR
UDI-DI
10811877011269
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Additional Manufacturer Narrative · 0

INVESTIGATION: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M161697 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT M161697 AND TEST BASE PART NUMBER 190-430 / LOT M161697. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M161697 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE AS THE LOGFILES WERE DELETED PRIOR TO EXPORT. A REVIEW OF COMPLAINTS AGAINST THE KIT LOT FOR THE REPORTED ISSUE INDICATES PRODUCT PERFORMANCE IS AS EXPECTED AND A PRODUCT DEFICIENCY HAS NOT BEEN IDENTIFIED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY FOR MULTIPLE PATIENTS PERFORMED ON (B)(6) 2021. THIS MFR. REPORT ADDRESSES PATIENT 1 OF 3. THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2021 ON A DIRECT TESTED NASAL KITTED SWAB. TESTING WAS PERFORMED AS ROUTINE PATIENT TESTING. CONFIRMATION TESTING WAS PERFORMED WITH AEGIS RAPID GENERATED NEGATIVE RESULTS. THE CUSTOMER STATED THE PATIENT WAS ASYMPTOMATIC. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1450102 ID NOW COVID-19 MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCABOROUGH M161697 10811877011269

Patients

Seq Age Sex Outcome Treatment
1 Unknown