FDA Adverse Event Malfunction Summary report: N

NDEHP PLUMSET 2CLAVES-SL

MDR report key: 12550314 · Received September 29, 2021

Report

Report Number
9615050-2021-00138
Event Type
Malfunction
Date Received
September 29, 2021
Date of Event
September 6, 2021
Report Date
September 6, 2021
Manufacturer
ICU MEDICAL COSTA RICA LTD.
Product Code
FPA
UDI-DI
10887787005650
PMA / PMN Number
K141789
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

RECEIVED AN IMAGE FROM THE CUSTOMER SHOWING THE SET, WITH FLUID, INSIDE OF A BAG. NO DAMAGE OR ANOMALIES WERE EVIDENT FROM THE IMAGE. RECEIVED ONE USED, LIST #146870489 (PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH; LOT #5192799) AND ONE USED NEEDLE (LIST #UNKNOWN, LOT #UNKNOWN). THE SET WAS VISUALLY EXAMINED AND A LARGE CUT WAS FOUND IN THE TUBING. THE SET WAS AIR PRESSURE LEAK TESTED AND THE AFOREMENTIONED CUT WAS THE ONLY LEAK SITE. THE CUT WAS EXAMINED AND THE EDGES WERE CLEAN. NO PACKAGING WAS RETURNED FOR EVALUATION. THE COMPLAINT CAN BE CONFIRMED, THE PROBABLE CAUSE IS DUE TO CONTACT WITH A SHARP INSTRUMENT DURING USE. THE MANUFACTURING PROCESS WAS REVIEWED AND IT WAS DETERMINED THAT THIS TYPE OF DAMAGE WOULD NOT HAVE OCCURRED DURING MANUFACTURING. THERE ARE NO SHARP OBJECTS OR INSTRUMENTS ON THE MANUFACTURING FLOOR CAPABLE OF THE OBSERVED DAMAGE AS A PREVENTATIVE MEASURE. THE LOT HISTORY WAS REVIEWED AND NO NONCONFORMITIES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE REPORTED COMPLAINT.

Additional Manufacturer Narrative · 1

THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT BEEN RECEIVED.

Description of Event or Problem · 1

THE EVENT INVOLVED A PLUM SET WHERE A HOLE WAS FOUND ON THE TUBING WHICH RESULTED IN A LEAK, 10 MINUTES AFTER INFUSION OF NICARDIPINE, FOR HIGH BLOOD PRESSURE TREATMENT. THERE WAS NO UNPROTECTED EXPOSURE TO THE DRUG LEAK. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT HARM. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1448143 NDEHP PLUMSET 2CLAVES-SL SET, ADMINISTRATION, INTRAVASCULAR FPA ICU MEDICAL COSTA RICA LTD. 1468728 5192799 10887787005650

Patients

Seq Age Sex Outcome Treatment
1 Unknown NICARDIPINE, MFR UNK