AQUA SEAL CHEST DRAINAGE UNIT WITH ATS

MDR report key: 125491 · Received October 10, 1997

Report

Report Number
9611018-1997-00008
Event Type
Other
Date Received
October 10, 1997
Date of Event
September 4, 1997
Report Date
September 12, 1997
Manufacturer
SHERWOOD MEDICAL INDUSTRIES
Product Code
KDQ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE LOTS IN STOCK ARE 701959, 97E213T, 97E215T, 97E273T. SAMPLES FROM EACH LOT WERE RETURNED AND EVALUATED. ALL TUBING AND CONNECTOR SPECS WERE ACCEPTABLE. PULL OUT FORCE WAS FINE. NO PRODUCT PROBLEMS WERE IDENTIFIED. A "Y" CONNECTOR USED BY THE HOSP WAS RETURNED, MEASURED AND FOUND TO BE COMPATIBLE WITH THE CDU. THE USE OF TWO DIFFERENT TYPES OF CONNECTORS (ONLY ONE SUPPLIED BY SD&G) APPEARS TO HAVE CAUSED THE REPORTED PROBLEM. THE CUSTOMER WAS ADVISED TO ENSURE SNUG FIT WITH EACH CONNECTOR. USER ERROR APPEARS TO BE THE CAUSE OF THE REPORTED PROBLEM.

Description of Event or Problem · 1

DURING THE WEEK OF 09/04 TO 09/10 THEY HAD THREE INCIDENTS IN WHICH THE PATIENT TUBE FROM THE AQUA SEAL DISCONNECTED FROM THE THORACIC CATHETER. TWO OCCURRED WHEN THE PATIENTS WERE TRANSPORTED FROM THE OR TO ICU. ONE OCCURRED AFTER THE PATIENT HAD BEEN IN ICU FOR FIVE DAYS. THIS PATIENT DEVELOPED A PNEUMOTHORAX AFTER THE TUBE DISCONNECTED. THE PNEUMOTHORAX HAS SINCE RESOLVED ITSELF. IN ALL CASES A NEW AQUA SEAL WAS PUT ON THE PATIENTS. THE CUSTOMER REPORTS, THEY NORMALLY PUT IN TWO CHEST TUBES AND CUT OFF THE CONNECTOR ON THE CDU AND REPLACE IT WITH THEIR "Y" CONNECTOR. THE CHEST TUBES ARE CONNECTED TO THE "Y" CONNECTOR. IN THIS INSTANCE, THE PATIENT TUBE CONNECTOR WAS USED AND THE STAFF DID NOT REALIZE THE CONNECTOR IS SMALLER THAN THE "Y" CONNECTOR. THE PATIENT TUBE CONNECTOR ON THE CDU DISCONNECTED FROM THE CHEST TUBE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AQUA SEAL CHEST DRAINAGE UNIT WITH ATS CHEST DRAINAGE/AUTO TRANSFUSION KDQ SHERWOOD MEDICAL INDUSTRIES NA 701959,97E213T,97E215T,97E273T

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Required Intervention