FDA Adverse Event Malfunction Summary report: N

S/C TRIAL HANDLE

MDR report key: 12547401 · Received September 29, 2021

Report

Report Number
1818910-2021-21411
Event Type
Malfunction
Date Received
September 29, 2021
Date of Event
September 15, 2021
Report Date
September 16, 2021
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
LXH
UDI-DI
10603295084471
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY : EXAMINATION OF THE RETURNED INSTRUMENT CANNOT CONFIRM THE COMPLAINT. DEPUY-SYNTHES CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED, AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT : 1) QUANTITY MANUFACTURED: (B)(4). 2) DATE OF MANUFACTURE: 3/3/2020. 3) ANY ANOMALIES OR DEVIATIONS IDENTIFIED IN DHR: 2 QHF'S-1 AT ASSEMBLY 1 AT FINAL. 4) EXPIRY DATE: NONE. 5) IFU REFERENCE: IFU-0902-00-836 REV J. DEVICE HISTORY REVIEW : THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT #(B)(4). INVESTIGATION SUMMARY: EXAMINATION OF THE RETURNED INSTRUMENT CANNOT CONFIRM THE COMPLAINT. DEPUY-SYNTHES CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED, AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT: THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED. CORRECTED: H3.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # : (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # : (B)(4). INVESTIGATION SUMMARY : EXAMINATION OF THE RETURNED INSTRUMENT CANNOT CONFIRM THE COMPLAINT. DEPUY-SYNTHES CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED, AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT: THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED.

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

CUSTOMER IS NOT ABLE TO CLEAN THE INSTRUMENT BEHIND SPRING BALLS. NO ADVERSE PATIENT IMPACT, NO SURGICAL DELAY. SEEN IN CSSD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1447230 S/C TRIAL HANDLE HIP INSTRUMENTS : HANDLES LXH DEPUY ORTHOPAEDICS INC US 2055-12-000 PG296356 10603295084471

Patients

Seq Age Sex Outcome Treatment
1 Unknown