CERTAIN® TITANIUM LARGE HEXED SCREW
Report
- Report Number
- 0001038806-2021-01835
- Event Type
- Malfunction
- Date Received
- September 29, 2021
- Date of Event
- August 18, 2021
- Report Date
- February 22, 2022
- Manufacturer
- BIOMET 3I
- Product Code
- DZE
- UDI-DI
- 00844868006138
- PMA / PMN Number
- K972444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). ONE (1) CERTAIN® TITANIUM LARGE HEXED SCREW (ILRGHT) WAS RETURNED FOR INVESTIGATION. VISUAL EVALUATION OF THE AS RETURNED PRODUCT IDENTIFIED THE SCREW FRACTURED ACROSS THE THREADS. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1230880). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. LOT NUMBERS WERE INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1230880) FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS FRACTURE SCREW. BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR AND THE REPORTED EVENT WAS CONFIRMED.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). PMA/510(K) NUMBER K072642.
IT WAS REPORTED THAT THE PROSTHETIC SCREW WAS FRACTURED. THE DOCTOR ALSO CONFIRMED THAT THE DENTAL SURGICAL PROCEDURE WAS COMPLETED WITH ANOTHER SCREW AND THAT THE PATIENT DID NOT SUFFER ANY NEGATIVE IMPACT ON HIS HEALTH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1444020 | CERTAIN® TITANIUM LARGE HEXED SCREW | DENTAL SCREW | DZE | BIOMET 3I | ILRGHT | 1230880 | 00844868006138 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female |