BD BACTEC¿ PEDS PLUS¿/ F CULTURE VIALS (PLASTIC)
Report
- Report Number
- 2647876-2021-00133
- Event Type
- Malfunction
- Date Received
- September 29, 2021
- Date of Event
- September 2, 2021
- Report Date
- October 25, 2021
- Manufacturer
- BECTON DICKINSON CARIBE LTD.
- Product Code
- MDB
- UDI-DI
- 00382904420208
- PMA / PMN Number
- K151866
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6. INVESTIGATION: CUSTOMER REPORTED A FALSE POSITIVE DEFECT. BD WAS UNABLE TO REPRODUCE CUSTOMER EXPERIENCE WITH THE BACTEC PRODUCT. A FALSE POSITIVE RESPONSE WAS NOT OBSERVED WHEN RETENTION SAMPLES WERE TESTED. BATCH HISTORY RECORDS WERE REVIEWED, AND ALL TESTING WERE WITHIN SPECIFICATION FOR PRODUCT RELEASE. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. USERS ARE CAUTIONED IN THE PACKAGE INSERT UNDER LIMITATION OF THE PROCEDURE: ¿A GRAM-STAINED SMEAR FROM CULTURE MEDIUM MAY CONTAIN SMALL NUMBER OF NON-VIABLE ORGANISMS DERIVED FROM MEDIA CONSTITUENTS, STAINING REAGENTS, IMMERSION OIL, GLASS SLIDE AND SPECIMENS USED FOR INOCULATION. THERE ARE MANY FACTORS THAT CAN INFLUENCE THE FALSE POSITIVE RATE, INCLUDING BLOOD VOLUME, BLOOD CELL COUNTS, ENVIRONMENTAL FACTORS, AND MEDIA LOT TO LOT VARIATIONS. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW. NO CORRECTIVES ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATION PROCESS. CATALOG 442020 BATCH NO. 1061192 CUSTOMER REPORTED A FALSE POSITIVE DEFECT. BD WAS UNABLE TO REPRODUCE CUSTOMER EXPERIENCE WITH THE BACTEC PRODUCT. A FALSE POSITIVE RESPONSE WAS NOT OBSERVED WHEN RETENTION SAMPLES WERE TESTED. BATCH HISTORY RECORDS WERE REVIEWED, AND ALL TESTING WERE WITHIN SPECIFICATION FOR PRODUCT RELEASE. BATCH HAS BEEN PREVIOUSLY INVESTIGATED FOR THE REPORTED DEFECT. NO ADDITIONAL TESTING IS REQUIRED AT THIS TIME. USERS ARE CAUTIONED IN THE PACKAGE INSERT UNDER LIMITATION OF THE PROCEDURE: ¿A GRAM-STAINED SMEAR FROM CULTURE MEDIUM MAY CONTAIN SMALL NUMBER OF NON-VIABLE ORGANISMS DERIVED FROM MEDIA CONSTITUENTS, STAINING REAGENTS, IMMERSION OIL, GLASS SLIDE AND SPECIMENS USED FOR INOCULATION. THERE ARE MANY FACTORS THAT CAN INFLUENCE THE FALSE POSITIVE RATE, INCLUDING BLOOD VOLUME, BLOOD CELL COUNTS, ENVIRONMENTAL FACTORS, AND MEDIA LOT TO LOT VARIATIONS. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW. NO CORRECTIVES ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATION PROCESS. PLOT/LOG FILE REVIEW: 7/8 PLOTS SHOWED NOISY SIGNAL. ABRUPT SIGNAL CHANGES SUCH AS THESE CAN CAUSE FALSE POSITIVES AND HAVE BEEN ASSOCIATED WITH PHYSICAL, ELECTRICAL, OR ENVIRONMENTAL CONDITIONS/DISRUPTIONS OF THE INSTRUMENT. TASK 3830574 WAS OPENED TO EVALUATE THE INSTRUMENT.
IT WAS REPORTED THAT WHILE USING BD BACTEC¿ PEDS PLUS¿/ F CULTURE VIALS (PLASTIC) FALSE POSITIVE RESULTS WERE OBTAINED BY THE LABORATORY PERSONNEL. A CULTURE AND BACTERIOSCOPY WERE USED TO CONFIRM THE RESULTS AS FALSE POSITIVES. THE CUSTOMER STATED RESULTS WERE NOT REPORTED OUT SO THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE VALUES FOR THE MONTH OF JULY AND AUGUST REPRESENT (B)(4) OF THE TOTAL ROUTINE, WHICH IS WITHIN EXPECTATIONS"
INITIAL REPORTER PHONE NUMBER: (B)(6). A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 1061192, MEDICAL DEVICE EXPIRATION DATE: UNKNOWN, DEVICE MANUFACTURE DATE: UNKNOWN, MEDICAL DEVICE LOT #: 1061193, MEDICAL DEVICE EXPIRATION DATE: 2021-03-02, DEVICE MANUFACTURE DATE: 2021-12-31.
IT WAS REPORTED THAT WHILE USING BD BACTEC¿ PEDS PLUS¿/ F CULTURE VIALS (PLASTIC) FALSE POSITIVE RESULTS WERE OBTAINED BY THE LABORATORY PERSONNEL. A CULTURE AND BACTERIOSCOPY WERE USED TO CONFIRM THE RESULTS AS FALSE POSITIVES. THE CUSTOMER STATED RESULTS WERE NOT REPORTED OUT SO THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE VALUES FOR THE MONTH OF JULY AND AUGUST REPRESENT 0.94% AND 1.97% OF THE TOTAL ROUTINE, WHICH IS WITHIN EXPECTATIONS".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1447216 | BD BACTEC¿ PEDS PLUS¿/ F CULTURE VIALS (PLASTIC) | SYSTEM, BLOOD CULTURING | MDB | BECTON DICKINSON CARIBE LTD. | 442020 | SEE H.10. | 00382904420208 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |