FDA Adverse Event Malfunction Summary report: N

RENEW MINI ENDOCUT SCISSOR TIP, DISPOSABLE

MDR report key: 12542613 · Received September 28, 2021

Report

Report Number
1223422-2021-00007
Event Type
Malfunction
Date Received
September 28, 2021
Date of Event
September 13, 2021
Report Date
September 28, 2021
Manufacturer
MICROLINE SURGICAL INC.
Product Code
GEI
UDI-DI
00811099010012
PMA / PMN Number
K962119
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WAS DECONTAMINATED AND FORWARDED TO QUALITY ENGINEERING FOR EVALUATION THE SCISSOR AND THE DETACHED SHRINK TUBING WERE RETURNED FOR EXAMINATION THE SCISSOR ASSEMBLY WAS MISSING THE SHRINK TUBING. THE SCISSOR HAS NO VISUAL DAMAGE EXCEPT FOR THE SHRINK TUBE THAT WAS NOT ATTACHED. THE SCISSORS WERE TESTED USING A KNOWN GOOD HAND PIECE. THE SCISSOR FUNCTIONED CORRECTLY. THE SHRINK TUBE WAS DAMAGED A SPLIT (LENGTH WISE). THE DISTAL EDGE OF THE SHRINK TUBE SHOWS SIGNS OF BEING ROLLED AND CUT. THIS APPEARANCE OF THE DAMAGE INDICATES THAT THE SHRINK TUBE WAS CUT BY IMPACT WITH A RIGID OBJECT

Description of Event or Problem · 1

DURING A PROCEDURE, OUR CUSTOMER FOUND THAT THE PART WHICH SHOULD BE ADHERED CAME OFF. THE SURGEON IMMEDIATELY PICKED IT UP THUS THERE WAS NEITHER HEALTH DAMAGE NOR EXTENSION OF THE PROCEDURE. A SPARE INSTRUMENT WAS USED TO CONTINUE THE PROCEDURE.