FDA Adverse Event Injury Summary report: N

UNKNOWN

MDR report key: 12541001 · Received September 28, 2021

Report

Report Number
1820334-2021-02229
Event Type
Injury
Date Received
September 28, 2021
Date of Event
August 19, 2011
Report Date
August 10, 2022
Manufacturer
COOK INC
Product Code
FGE
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION ¿ EVALUATION. IN THE LITERATURE ARTICLE " TESTICULAR ARTERY EMBOLIZATION FOR THE TREATMENT OF IATROGENIC HEMORRHAGE: REPORT OF TWO CASES¿ IT WAS REPORTED THAT INJURY TO A TESTICULAR ARTERY OCCURRED POSSIBLY DUE TO A DRAINAGE CATHETER EXCHANGE. THE PATIENT REQUIRED AN UNKNOWN 12FR PIGTAIL DRAINAGE CATHETER FOR TREATMENT OF RETROPERITONEAL FLUID COLLECTION FOLLOWING DUODENAL PERFORATION AFTER ENDOSCOPIC ULTRASOUND AND BIOPSY. AFTER 4 DAYS OF CONTINUED HIGH-OUTPUT DRAINAGE, THE 10-F DRAINAGE CATHETER WAS REPLACED AND REPOSITIONED FOR BETTER CONTROL OF THE LEAK. OVER THE NEXT 2 DAYS, THE PATIENT HAD INTERMITTENT BLOODY OUTPUT, AND A CT SCAN SHOWED A POSSIBLE VASCULAR PSEUDOANEURYSM IN THE REGION OF THE DRAIN. A MESENTERIC ANGIOGRAM WAS REQUESTED 4 DAYS AFTER THE FIRST CATHETER EXCHANGE. INJECTION OF CONTRAST AGENT IN THE RIGHT RENAL ARTERY SHOWED AN AREA OF ACTIVE EXTRAVASATION IMMEDIATELY ADJACENT TO THE DRAINAGE CATHETER. THE TESTICULAR ARTERY WAS SELECTIVELY INJECTED USING A MICROCATHETER AND EMBOLIZATION WAS PERFORMED BOTH PROXIMAL AND DISTAL TO THE PSEUDOANEURYSM. NO FURTHER ACTIVE BLEEDING WAS SEEN AFTER EMBOLIZATION. NO OTHER ADVERSE EVENTS WERE REPORTED FOR THIS EVENT REVIEWS OF DOCUMENTATION INCLUDING THE QUALITY CONTROL PROCEDURES AND INSTRUCTIONS FOR USE (IFU) WERE CONDUCTED DURING THE INVESTIGATION. THE COMPLAINT DEVICE WAS NOT RETURNED; THEREFORE, NO PHYSICAL EXAMINATIONS COULD BE PERFORMED. HOWEVER, A DOCUMENT-BASED INVESTIGATION EVALUATION WAS PERFORMED. A REVIEW OF THE DEVICE MASTER RECORD (DMR) CONCLUDED THAT SUFFICIENT INSPECTION ACTIVITIES ARE IN PLACE TO IDENTIFY THIS FAILURE MODE PRIOR TO DISTRIBUTION. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS UNABLE TO BE COMPLETED DUE TO A LACK OF LOT INFORMATION. COOK REVIEWED THE SALES HISTORY FOR THIS CUSTOMER AND WAS UNABLE TO IDENTIFY THE COMPLAINT LOT. BASED ON THE AVAILABLE INFORMATION, COOK HAS CONCLUDED THAT THE DEVICE WAS MANUFACTURED TO SPECIFICATION AND THAT THERE IS NO EVIDENCE SUGGESTING NONCONFORMING PRODUCT EXISTS EITHER IN HOUSE OR IN FIELD. COOK ALSO REVIEWED PRODUCT LABELING. THE PRODUCT IFU, [T_MULTI_REV5] ¿MULTIPURPOSE DRAINAGE CATHETER,¿ PROVIDES THE FOLLOWING INFORMATION TO THE USER RELATED TO THE REPORTED FAILURE MODE: ¿INTENDED USE MULTIPURPOSE DRAINAGE CATHETERS ARE INTENDED FOR PERCUTANEOUS DRAINAGE IN A VARIETY OF DRAINAGE APPLICATIONS (E.G., NEPHROSTOMY, BILIARY, AND ABSCESS), EITHER BY DIRECT STICK OR SELDINGER ACCESS TECHNIQUE. HOW SUPPLIED: SUPPLIED STERILIZED BY ETHYLENE OXIDE GAS IN PEEL-OPEN PACKAGES. INTENDED FOR ONE-TIME USE. STERILE IF PACKAGE IS UNOPENED OR UNDAMAGED. DO NOT USE THE PRODUCT IF THERE IS DOUBT AS TO WHETHER THE PRODUCT IS STERILE. STORE IN A DARK, DRY, COOL PLACE. AVOID EXTENDED EXPOSURE TO LIGHT. UPON REMOVAL FROM PACKAGE, INSPECT THE PRODUCT TO ENSURE NO DAMAGE HAS OCCURRED.¿ BASED ON THE AVAILABLE INFORMATION, NO DEVICE RETURN, AND THE RESULTS OF THE INVESTIGATION, COOK CONCLUDED THE CAUSE OF THE EVENT TO BE ADVERSE EVENT RELATED TO PROCEDURE. IT WAS REPORTED THAT INJURY TO A TESTICULAR ARTERY OCCURRED POSSIBLY DUE TO A DRAINAGE CATHETER EXCHANGE. IT IS POSSIBLE DIFFICULTY ADVANCING THE CATHETER MAY HAVE LED THE USER TO USE EXCESSIVE FORCE WHICH COULD HAVE CONTRIBUTED TO THE EVENT. THIS CANNOT BE CONFIRMED WITHOUT ADDITIONAL INFORMATION. THE APPROPRIATE PERSONNEL HAVE BEEN NOTIFIED. COOK WILL CONTINUE TO MONITOR FOR SIMILAR COMPLAINTS. PER THE RISK ASSESSMENT NO FURTHER ACTION IS REQUIRED. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803 AND IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT CONTAINED HEREIN IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, NOR THAT ANY COOK DEVICE CAUSED, CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

Additional Manufacturer Narrative · 1

BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN OR UNAVAILABLE. SUSPECT MEDICAL DEVICE: DEVICE IS AN UNSPECIFIED 12.0 FR PIGTAIL DRAINAGE CATHETER. COMMON DEVICE NAME: UNKNOWN AS EXACT DEVICE IDENTITY IS NOT KNOWN. PMA/510(K) #: UNKNOWN AS EXACT DEVICE IDENTITY IS NOT KNOWN. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE.

Description of Event or Problem · 1

THE BELOW REPORT OF A TESTICULAR ARTERY INJURY POSSIBLY DUE TO DRAINAGE CATHETER EXCHANGE WAS TAKEN FROM THE FOLLOWING ARTICLE: NIEDERHAUSER,B. ET. AL., TESTICULAR ARTERY EMBOLIZATION FOR THE TREATMENT OF IATROGENIC HEMORRHAGE: REPORT OF TWO CASES. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY 22.9 (2011): 1347-1348. THE ARTICLE CONTAINS "TWO CASES OF TESTICULAR ARTERY EMBOLIZATION PERFORMED FOR THE TREATMENT OF IATROGENIC HEMORRHAGE...[ONE] WAS OF UNCERTAIN ETIOLOGYRESULTING EITHER FROM AN ENDOSCOPIC DUODENAL PERFORATION OR FROM INJURY DURING A DRAINAGE CATHETER EXCHANGE (CASE REPORT 1)." A (B)(6) MAN WITH UNRESECTABLE PANCREATIC ADENOCARCINOMA PRESENTED TO OUR DEPARTMENT WITH A DUODENAL PERFORATION AFTER ENDOSCOPIC ULTRASOUND AND BIOPSY. A RESULTING RETROPERITONEAL FLUID COLLECTION WAS CONTROLLED USING A 12.0-F PIGTAIL DRAINAGE CATHETER (COOK, INC, BLOOMINGTON, INDIANA) PLACED UNDER COMPUTED TOMOGRAPHY (CT) GUIDANCE FROM A LATERAL APPROACH USING A MODIFIED SELDINGER TECHNIQUE. AFTER 4 DAYS OF CONTINUED HIGH-OUTPUT DRAINAGE, A SINOGRAM WAS PERFORMED, WHICH SHOWED COMMUNICATION TO THE THIRD PORTION OF THE DUODENUM THROUGH A SIZABLE (1 CM) DEFECT. THE 10-F DRAINAGE CATHETER (MULTIPURPOSE DRAINAGE CATHETER; COOK, INC) WAS REPLACED OVER A STIFF GUIDE WIRE (LUNDERQUIST-RING TORQUE WIRE, COOK, INC) AND REPOSITIONED FOR BETTER CONTROL OF THE LEAK. OVER THE NEXT 2 DAYS, THE PATIENT HAD INTERMITTENT BLOODY OUTPUT FROM THE DRAINAGE CATHETER AND NASOGASTRIC TUBE, AND A CT SCAN SHOWED A POSSIBLE VASCULAR PSEUDOANEURYSM IN THE REGION OF THE DRAIN WITH NO SIGNIFICANT RETROPERITONEAL HEMATOMA. ENDOSCOPY WITH CLIPPING OF THE DUODENAL PERFORATION WAS PERFORMED; NO ENDOLUMINAL BLOOD WAS NOTED. A MESENTERIC ANGIOGRAM WAS REQUESTED 4 DAYS AFTER THE FIRST CATHETER EXCHANGE WHEN FRANK BLOOD AND CLOT WERE AGAIN NOTED IN THE NASOGASTRIC TUBE AND DRAINAGE CATHETER, ALONG WITH A 4-DAY HEMOGLOBIN DECLINE OF 4.3 G/DL (12.9 G/DL TO 8.6 G/DL). INJECTION OF CONTRAST AGENT IN THE RIGHT RENAL ARTERY SHOWED THE RIGHT TESTICULAR ARTERY ARISING FROM THE UPPER POLE BRANCH OF THE RIGHT RENAL ARTERY WITH AN AREA OF ACTIVE EXTRAVASATION IMMEDIATELY ADJACENT TO THE DRAINAGE CATHETER (FIG, A). THE TESTICULAR ARTERY WAS SELECTIVELY INJECTED USING A MICROCATHETER. EMBOLIZATION WAS PERFORMED BOTH PROXIMAL AND DISTAL TO THE PSEUDOANEURYSM USING TWO 0.018-INCH, 1.0 CM LONG (HILAL EMBOLIZATION MICROCOIL; COOK, INC) AND TWO 0.018-INCH 3 X 2 MM TORNADO (COOK, INC) COILS; NO FURTHER ACTIVE BLEEDING WAS SEEN AFTER EMBOLIZATION. NO OTHER ADVERSE EFFECTS WERE REPORTED OR CONTAINED WITHIN THE ARTICLE. THE 12 FR PIGTAIL DRAIN IS CAPTURED IN THE REPORT WITH PATIENT IDENTIFIER (B)(6). THE 10 FR MULTIPURPOSE DRAIN IS CAPTURED IN THE REPORT WITH PATIENT IDENTIFIER (B)(6). THE LUNDERQUIST-RING TORQUE WIRE IS CAPTURED IN THE REPORT WITH PATIENT IDENTIFIER (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1441846 UNKNOWN FGE COOK INC N/A UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 73 YR Male Required Intervention COOK 10 FR MULTIPURPOSE DRAIN| COOK LUNDERQUIST-RING TORQUE WIRE