FDA Adverse Event Malfunction Summary report: N

ARCHITECT STAT HIGH SENSITIVE TROPONIN-I

MDR report key: 12540847 · Received September 28, 2021

Report

Report Number
3005094123-2021-00185
Event Type
Malfunction
Date Received
September 28, 2021
Date of Event
September 12, 2021
Report Date
November 22, 2021
Manufacturer
A.I.D.D LONGFORD
Product Code
MMI
PMA / PMN Number
K191595
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE LOT SEARCH REVIEW DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY FOR THE ISSUE FOR THE LOT. A REVIEW OF TRACKING AND TRENDING DID NOT IDENTIFY ANY TRENDS FOR THE COMPLAINT ISSUE. RETURN TESTING WAS NOT PERFORMED AS RETURNS WERE NOT AVAILABLE. DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY ISSUES ASSOCIATED WITH LOT NUMBER 30254UD00 AND THE COMPLAINT ISSUE. THE PERFORMANCE OF THE ARCHITECT STAT HIGH SENSITIVE TROPONIN-I ASSAY WAS ASSESSED USING WORLD-WIDE FIELD DATA AND SHOWS THAT THE MEDIAN PATIENT RESULT FOR LOT 30254UD00 IS WITHIN ESTABLISHED LIMITS AND COMPARABLE WITH OTHER LOTS IN THE FIELD, CONFIRMING NO SYSTEMIC ISSUE FOR THE LOTS. LABELING WAS REVIEWED WHICH ADEQUATELY ADDRESSES THE CURRENT ISSUE. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ARCHITECT STAT HIGH SENSITIVE TROPONIN-I REAGENT LOT 30254UD00 WAS IDENTIFIED.

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 03P25 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 02R98. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT A PHYSICIAN WAS QUESTIONING FALSELY ELEVATED ARCHITECT STAT HIGH SENSITIVE TROPONIN-I RESULTS FOR ONE FEMALE PATIENT. THE FOLLOWING DATA WAS PROVIDED (FEMALE REFERENCE RANGE <16 PG/ML): ON (B)(6) 2021 SAMPLE (B)(6). INITIAL RESULT WAS 136 PG/ML, REPEATS 98 / 584 / 1745 / <5 / <5 PG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1437083 ARCHITECT STAT HIGH SENSITIVE TROPONIN-I IMMUNOASSAY METHOD, TROPONIN SUBUNIT MMI A.I.D.D LONGFORD 30254UD00

Patients

Seq Age Sex Outcome Treatment
1 Female ARC I2000SR INST, 03M74-02, (B)(6)| ARC I2000SR INST, 03M74-02, (B)(6)| ARC I2000SR INST, 03M74-02, ISR07891