HUDSON AIR CUSHION MASK,NEONATE
Report
- Report Number
- 3011137372-2021-00248
- Event Type
- Malfunction
- Date Received
- September 27, 2021
- Date of Event
- July 20, 2020
- Report Date
- September 1, 2021
- Product Code
- BSJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
(B)(4). COMPLAINT VERIFICATION TESTING COULD NOT BE PERFORMED AS IT WAS REPORTED THAT THE SAMPLE IS NOT AVAILABLE FOR RETURN. THE MANUFACTURER REPORTS THAT THE MANUFACTURING AND INSPECTION RECORDS OF LOT#180618 WERE REVIEWED AND NO ISSUES RELATED TO THE REPORTED COMPLAINT WERE FOUND. THE MANUFACTURER ALSO REPORTS THAT "THE MASK IS MADE OF PVC MATERIAL THROUGH BLOW MOLDING, THE MOLECULAR WEIGHT AND THE MOLECULAR DENSITY OF THE MASK IS HIGH, THE CUSHION WILL DEFLATE NATURALLY AT THE SEALING JOINT AFTER LONG TERM STORAGE." WITHOUT THE DEVICE TO EVALUATE THE COMPLAINT COULD NOT BE CONFIRMED AND THE PROBABLE CAUSE COULD NOT BE DETERMINED FROM THE AVAILABLE INFORMATION. TELEFLEX WILL CONTINUE TO MONITOR AND TREND FOR REPORTS OF THIS NATURE. CORRECTED DATA: SECTION B.4.-DATE OF THIS REPORT CORRECTED TO (B)(6) 2021.
IT WAS REPORTED: "EXTREMELY PREMATURE BABY OF 25 WEEKS OF GESTATION, LOW WEIGHT FOR THE GESTATIONAL AGE WITH NEWBORN RESPIRATORY DISTRESS SYNDROME NEEDED INTUBATION , AN AIR CUSHION MASK WAS REQUIRED, ALTERATION IN THE PRESSURE OF THE CUSHION FILLING (DEFLATION) WAS OBSERVED AND THE DOCTOR REQUESTED TO CHANGE THE MASK. PATIENT AGE: 0. PATIENT HEIGHT: 35CM. PATIENT WEIGHT: 0.725KG. ADDITIONAL INFORMATION RECEIVED INDICATES "THE INCIDENT HAPPENED LAST YEAR, INFORMATION ABOUT THE PATIENT WAS NOT PROVIDED. THE ISSUE IN THE PRODUCT WAS DETECTED PRIOR TO USE, SO THE DOCTOR JUST CHANGED THE MASK." NO PATIENT INJURY OR HARM REPORTED.
(B)(4).
IT WAS REPORTED: "EXTREMELY PREMATURE BABY OF 25 WEEKS OF GESTATION, LOW WEIGHT FOR THE GESTATIONAL AGE WITH NEWBORN RESPIRATORY DISTRESS SYNDROME NEEDED INTUBATION , AN AIR CUSHION MASK WAS REQUIRED, ALTERATION IN THE PRESSURE OF THE CUSHION FILLING (DEFLATION) WAS OBSERVED AND THE DOCTOR REQUESTED TO CHANGE THE MASK. PATIENT AGE: (B)(6), PATIENT HEIGHT: 35CM, PATIENT WEIGHT: (B)(6)KG. ADDITIONAL INFORMATION RECEIVED INDICATES "THE INCIDENT HAPPENED LAST YEAR, INFORMATION ABOUT THE PATIENT WAS NOT PROVIDED. THE ISSUE IN THE PRODUCT WAS DETECTED PRIOR TO USE, SO THE DOCTOR JUST CHANGED THE MASK." NO PATIENT INJURY OR HARM REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1431007 | HUDSON AIR CUSHION MASK,NEONATE | ANESTHETIC GAS MASK | BSJ | 180618 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |