BD LUER-LOK TIP SYRINGE
Report
- Report Number
- 1911916-2021-01023
- Event Type
- Malfunction
- Date Received
- September 27, 2021
- Date of Event
- July 22, 2021
- Report Date
- September 10, 2021
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- FMF
- UDI-DI
- 30382903096535
- PMA / PMN Number
- K110771
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: IT WAS REPORTED WHEN PREPARING MEDICATION A BD LUER-LOK SYRINGE HAS A SMALL PIECE OF PLASTIC IN THE TIP OF THE SYRINGE. TO AID IN THE INVESTIGATION, TWO PHOTOS WERE PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. BOTH PHOTOS SHOW A SYRINGE WITH A LUER TIP THAT HAS AN EXCESS OF PLASTIC. THE SYRINGE BARREL SHOWS MARKS ON ONE SIDE THAT GO UP TO THE LUER TIP AS WELL. NO OTHER DEFECTS OR IMPERFECTIONS WERE OBSERVED. THIS DEFECT COULD OCCUR DURING THE MOLDING PROCESS IF THE TEMPERATURE DROPPED WHILE THE PART WAS BEING MOLDED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 309653, LOT NUMBER 0062293. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. MOLDING PARAMETERS WERE VERIFIED FINDING THEM ACCEPTABLE. TO DATE, THERE HAVE BEEN NO OTHER SIMILAR EVENTS REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND WITH THE PHOTO SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT A PIECE OF PLASTIC WAS FOUND IN THE BD LUER-LOK¿ TIP SYRINGE TIP WHILE PREPARING MEDICINE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "FOREIGN OBJECT INSIDE SYRINGE TIP WHEN PREPARING MEDICATION, IT IS DISCOVERED THAT A BD LUER-LOK SYRINGE HAS A SMALL PIECE OF PLASTIC IN THE TIP OF THE SYRINGE."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1430344 | BD LUER-LOK TIP SYRINGE | PISTON SYRINGE | FMF | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 309653 | 0062293 | 30382903096535 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |