FDA Adverse Event Injury Summary report: N

PUMP MMT-722RNAL PRDGM INS CL EN RC

MDR report key: 1253358 · Received December 2, 2008

Report

Report Number
2032227-2008-02005
Event Type
Injury
Date Received
December 2, 2008
Date of Event
November 11, 2008
Report Date
November 11, 2008
Manufacturer
MEDTRONIC MINIMED
Product Code
LZG
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE CUSTOMER'S DAUGHTER REPORTED THAT THE CUSTOMER WAS TREATED BY PARAMEDICS DUE TO HYPOGLYCEMIA. THE REPORTED BLOOD GLUCOSE READING WAS 32 MG/DL. TROUBLESHOOTING WAS PERFORMED AND FOUND THAT THE INSULIN PUMP WAS PROGRAMMED AND READING THE RESERVOIR VOLUME CORRECTLY. THE DISPLACEMENT TEST PASSED. WHEN THE HISTORY FILES OF THE INSULIN PUMP WERE CHECKED, IT WAS FOUND THAT THE BOLUS AND BASAL TOTALS DID NOT ADD UP CORRECTLY. THE CUSTOMER WAS ADVISED TO REVERT TO A BACKUP PLAN TO TREAT HER DIABETES. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722RNAL PRDGM INS CL EN RC INFUSION PUMP LZG MEDTRONIC MINIMED MMT-722RNAL

Patients

Seq Age Sex Outcome Treatment
1 58 YR Required Intervention