M530 OHX
Report
- Report Number
- 3003974370-2021-00017
- Event Type
- Malfunction
- Date Received
- September 24, 2021
- Date of Event
- September 14, 2021
- Report Date
- October 1, 2021
- Manufacturer
- LEICA MICROSYSTEMS (SCHWEIZ) AG
- Product Code
- EPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
THIS IS A FINAL REPORT. AN INVESTIGATION OF THE REPORTED INCIDENT WAS CONDUCTED AND REVEALED, THAT THE MALFUNCTION WAS DUE TO A BURNT NTC (R15) ON THE VOLTAGE SELECTOR BOARD. TESTS ON A REPRESENTATIVE M530 OHX, A DESIGN REVIEW OF THE ELECTRONIC CIRCUITS INCLUDING COMPONENT SPECIFICATIONS AND AN INVESTIGATION ON THE AFFECTED VOLTAGE SELECTOR BOARD WAS PERFORMED. THE TESTS AND INVESTIGATION REVEALED THAT THE DESIGN IS ADEQUATE AND THE NTC "R15" IS WORKING WITHIN ITS SPECIFIED LIMITS. IT WAS CONCLUDED THAT THE DEFECTIVE COMPONENT NTC "R15" IS AN ISOLATED, RANDOM COMPONENT FAILURE WITHOUT ANY DESIGN OR MANUFACTURING ISSUE AND NO INDICATIONS FOR AN ADVERSE TREND. REPLACING THE VOLTAGE SELECTOR BOARD CORRECTED THE ISSUE AND THE DEVICE WAS FUNCTIONING ACCORDING TO SPECIFICATIONS.
AN INVESTIGATION OF THE INCIDENT IS CURRENTLY UNDERWAY AND A FOLLOW-UP WILL BE SUBMITTED SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE FOLLOWING INVESTIGATION.
LEICA MICROSYSTEMS (B)(4) RECEIVED A COMPLAINT FROM USA STATING THAT AN M530 OHX UNEXPECTEDLY SHUT OFF DURING A PROCEDURE. THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1422639 | M530 OHX | SURGICAL MICROSCOPE | EPT | LEICA MICROSYSTEMS (SCHWEIZ) AG | M530 OHX |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |