DEPTH GAUGE FOR MULTILOC HUMERAL NAILING SYSTEM
Report
- Report Number
- 8030965-2021-07939
- Event Type
- Malfunction
- Date Received
- September 23, 2021
- Date of Event
- August 25, 2021
- Report Date
- August 25, 2021
- Manufacturer
- SYNTHES GMBH
- Product Code
- HTJ
- UDI-DI
- 07611819427174
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: D9 - DATE DEVICE RETURNED TO MANUFACTURER. H6 - CODES UPDATED TO IMDRF CODES. THE PRODUCT WAS RETURNED TO US CQ FOR EVALUATION. THE US CQ TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT DEPTH GAUGE F/ML HUM NAIL THE DEPTH GAUGE TIP WAS BROKEN, AND BROKEN PIECE WAS ALSO RETURNED NO OTHER ISSUES WERE IDENTIFIED. THE DIMENSIONAL INSPECTION WAS NOT PERFORMED DUE TO POST MANUFACTURING DAMAGE. THE OBSERVED CONDITION DEPTH GAUGE F/ML HUM NAIL IN THE DEVICE WAS CONSISTENT WITH THE COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED BROKEN CONDITION OF THE DEPTH GAUGE F/ML HUM NAIL. WHILE NO DEFINITIVE ROOT CAUSE COULD BE DETERMINED. THERE WAS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE CONTRIBUTED TO THE COMPLAINT. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DRAWING/SPECIFICATIONS REVIEWED. THE FOLLOWING DRAWING REFLECTING THE CURRENT AND MANUFACTURE REVISION WAS REVIEWED: - DEPTH GAUGE EXPERT- SE_408386 REV D/C. DEVICE HISTORY LOT - PART: 03.019.017. LOT: L381819. MANUFACTURING SITE: HAGENDORF. RELEASE TO WAREHOUSE DATE: 09 JUNE 2017. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
INITIAL REPORTER IS A SYNTHES EMPLOYEE THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, IT WAS REPORTED THAT THE DEPTH GAUGE WAS RECEIVED WITH THE TIP BROKEN. ANOTHER DEPTH GAUGE WAS AVAILABLE FOR USE IN THE PROCEDURE. THIS REPORT IS FOR A DEPTH GAUGE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1415638 | DEPTH GAUGE FOR MULTILOC HUMERAL NAILING SYSTEM | GAUGE, DEPTH | HTJ | SYNTHES GMBH | L381819 | 07611819427174 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |