FDA Adverse Event Injury Summary report: N

STARCLOSE SE VASCULAR CLOSURE SYSTEM

MDR report key: 1251295 · Received December 3, 2008

Report

Report Number
2953144-2008-01949
Event Type
Injury
Date Received
December 3, 2008
Date of Event
November 6, 2008
Report Date
November 10, 2008
Manufacturer
ABBOTT VASCULAR - REDWOOD CITY
Product Code
MGB
PMA / PMN Number
P050007
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. THE LOT NUMBER WAS NOT IDENTIFIED; THEREFORE, A DEVICE HISTORY RECORD REVIEW COULD NOT BE PERFORMED.

Description of Event or Problem · 1

DEVICE MALFUNCTION: DIFFICULT TO REMOVE. TIME OF DEVICE MALFUNCTION: DURING VESSEL CLOSURE. SYMPTOMS/AE: FAILURE TO ACHIEVE HEMOSTASIS. IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE STARCLOSE SE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER AN UNSPECIFIED PROCEDURE. REPORTEDLY, AFTER CLIP DEPLOYMENT THE DEVICE WAS DIFFICULT TO REMOVE. AN ATTEMPT WAS MADE TO REMOVE THE DEVICE WITH THE DILATOR-ASSISTED PORTS BUT WAS UNSUCCESSFUL. THE DEVICE WAS REMOVED USING COUNTER-TRACTION AND AN ASSERTIVE PULL. MANUAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STARCLOSE SE VASCULAR CLOSURE SYSTEM MGB ABBOTT VASCULAR - REDWOOD CITY NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention